www.vtiinstruments.com
EX1401 Instrument Drivers
89
Object
Name String
Value
Cold
Junction
Temperature
Sensor Reading
Reference Junction
The temperature of the CJC sensor, in the same
units as the channel data.
timestamp_sec
The seconds portion of the timestamp of the
first TC sample that this CJC measurement was
used to compute.
timestamp_frac
The fractional portion of the timestamp of the
first TC sample that this CJC measurement was
used to compute.
Over range
Over-Range
True when any overload status bit is set.
Trigger time
Trigger Timestamp
Seconds
The trigger timestamp in seconds
Trigger Timestamp
Fraction
The fraction portion of the trigger timestamp.
Dropped trigger
timestamp_sec
The timestamp of when the trigger would have
occurred.
timestamp_frac
The fraction portion.
Trigger Dropped
The channel name of the lost trigger.
T
ABLE
5-4:
A
DDITIONAL
D
ATA
JSON
N
AMES AND
V
ALUES
M
EASUREMENT
P
ROCESS
M
EASUREMENT
S
ETUP
When the driver session is initialized by the
Initialize
method, the driver session is reset, or after the
previous measurement is finished, the instrument is in the
Idle
state. While in the
Idle
state, the user
prepares for the next data acquisition by configuring the setup parameters. For more information,
please refer to IVILXISync standard API documentation
Sync and Coordination
The
Sync
interface controls how the EX1401 synchronizes its ADCs and its timestamp counter.
When a sync event occurs, all ADCs in the system are synchronized to each-other, and all data
timestamp counters are reset and, if using the System Timestamp Source, synchronized to the IEEE
1588 clock. If the
Sync.CoordinationLine
property is set to “LAN”, all instruments in the driver
session will use LXI Lan Event messages to coordinate their trigger state machines, ensuring that
all arm and trigger events are shared between all instruments in the system.
Start, Arm, and Trigger
Properties in the
Start
,
Arm
, and
Trigger
interfaces configure the data acquisition gating condition
and timing.
Start
determines when the data acquisition starts and when the instrument is ready for
the next acquisition. All hardware configuration and filters settling should be completed before the
Start
event in order to ensure valid data is acquired.
Sampling Parameters
The sampling parameters are specified by properties in
Sampling
interface.
Sampling
in the
Measurement
interface configures parameters common to the entire system. Optionally, the user can
configure the parameters of individual input channels using the
Channels
interface (with some
restrictions).
Содержание EX1401
Страница 2: ......
Страница 24: ......
Страница 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Страница 94: ......
Страница 98: ......
Страница 102: ......
Страница 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Страница 112: ......
Страница 114: ...VTI Instruments Corp 114 EX1401 Index ...