VTI Instruments Corp.
104
EX1401 Calibration
O
BTAIN
O
FFSET COEFFICIENTS
This procedure is performed to evaluate the offset error on each channel path. In order to perform
this step, you will need to build a special custom cable assembly that connects the thermocouple
simulator to 16 uncompensated thermocouple plugs, as shown in
Error! Reference source not
found.
1.
ECTRON 1140A
EX1401
Ectron
16x
F
IGURE
8-1:
G
AIN
E
RROR
M
EASUREMENT
S
ETUP
1.
Attach the Calibration cable between the Ectron calibrator and all channels of the EX1401.
a.
Red banana plug of TC Simulator cable plugs into ‘+’ on Ectron TC Simulator
b.
Black banana plug of TC Simulator cable plugs into ‘-‘ on Ectron
TC Simulator
c.
Secure earth ground to the attachment points on the Ectron TC Simulator and
EX1401
2.
In the Soft Front Panel ‘Channels’ table, enable all channels and configure them to Voltage
function, 10V range and 10 Record Size as shown in
Error! Reference source not found.
2.
F
IGURE
8-2:
U
SER
F
ULL
C
ALIBRATION WEB PAGE
3.
Ensure that Clock is set to 10000 and Sample Rate is set to 9.765625 in the ‘Sampling’ table,
as shown in
Error! Reference source not found.
3.
F
IGURE
8-3:
S
AMPLING
W
INDOW
4.
Set Records Per Trigger to 1 in the ‘Measurement’ table. Set all other fields as shown in
Error!
Reference source not found.
4.
Содержание EX1401
Страница 2: ......
Страница 24: ......
Страница 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Страница 94: ......
Страница 98: ......
Страница 102: ......
Страница 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Страница 112: ......
Страница 114: ...VTI Instruments Corp 114 EX1401 Index ...