VTI Instruments Corp.
80
EX1401 Instrument Drivers
R
EFERENCE
C
LOCK AND
T
IME
S
TAMPS
The
Source
property and
TimestampSource
property in the
ReferenceOscillator
interface configure
the reference oscillator to generate an ADC sampling clock and data time stamp clock. The
PXIe_CLK100 (100 MHz) reference signal is used by default. When the system is synchronized to
the IEEE 1588 PTP grand master via Ethernet, the IEEE 1588 clock can be used as a reference
oscillator be selecting “System”.
LXI
AND
LAN
E
VENTS
The EX1401 can act as an LXI device. These instruments can be synchronized to an IEEE 1588
PTP grand master clock. They can be armed or triggered by LAN events or they can generate LXI
LAN events to synchronize with other LXI devices through the IVI-LXISync interface defined in
the IVI driver specifications.
F
IGURE
5-6:
LXI
AND
LAN
E
VENTS
IVI
C
LASS
C
OMPLIANT
I
NTERFACES
In addition to the IVI-LXISync interface, the VTEXDigitizer/VTEXDsa drivers expose the IVI-
Digitizer and IVI-Scope class compliant interfaces. These interfaces are exposed for the instruments
interchangeability, but their use is not recommended. These IVI interfaces do not support the full
capabilities of the EX1401’s sophisticated trigger model and many other features.
Содержание EX1401
Страница 2: ......
Страница 24: ......
Страница 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Страница 94: ......
Страница 98: ......
Страница 102: ......
Страница 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Страница 112: ......
Страница 114: ...VTI Instruments Corp 114 EX1401 Index ...