Appendix
90
EzRay Air Portable (Model: VEX-P300) User Manual
A.4.3
Scattered Dose
Scope
IEC 60601-2-65 203.13
Requirements
ME EQUIPMENT shall be provided with means to optionally allow actuation of the
EXPOSURE from a PROTECTED AREA after installation.
Relevant instructions shall be given in the ACCOMPANYING DOCUMENTS.
Results
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively.
Test Equipment
Instrument
Manufacturer
Model
S/N
X / Gamma
Survey Meter
Radcal Co.
9015/10X5-180
91-1470/19069
Method
PMMA Phantom aligned to 280 mm away from Focal Spot
Max. Exposure Condition
Measure point: 1000 mm from PMMA Phantom
Содержание EzRay Air Portable
Страница 2: ......
Страница 4: ...Notice ii EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 24: ...3 System Overview 22 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 48: ...4 Operation 46 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 64: ...5 Service Mode 62 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 72: ...8 Disposing of the Unit 70 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...
Страница 76: ...9 Device Specifications 74 EzRay Air Portable Model VEX P300 User Manual 4 Tube Dimensions mm ...
Страница 101: ...Appendix EzRay Air Portable Model VEX P300 User Manual 99 ENGLISH This page is intentionally left blank ...
Страница 102: ...Appendix 100 EzRay Air Portable Model VEX P300 User Manual This page is intentionally left blank ...