© 2017 Van Putten Instruments BV | MAN-VP-SINL-UK | Revision:1701 | Date:22-03-2017
29
11
Specifications
Please always check the label of your product for the specifications.
Specifications are subject to change as we are continuously improving our products.
Please contact us to obtain the latest specification sheet.
Flow sensor
(minimum detection level and max flow rate shown)
- VPS.R080.M050
0.23...80 m
3
n
/hr
0.13…50 SCFM
- VPS.R250.M100
0.91...250 m
3
n
/hr
0.54…150 SCFM
- VPS.R01K.M200
3.55...1000 m
3
n
/hr
2.15…600 SCFM
Accuracy
0.5% FSS under calibration conditions with calibration report
5% FSS without calibration report
Reference conditions
0°C, 1013.25 mbar - DIN1343
32°F, 14.65 psi
Gas temperature
0....60°C
32..140°F
Gases
Compressed air, nitrogen, inert gases,
95% non condensing gases
Pressure sensor
Range
0...16 bar gauge
0…250 psi gauge
Accuracy
+/- 1.5% FSS (0...60°C)
+/- 1.5% FSS (32..140°F)
Temperature sensor
Range
0....60°C
32…140°F
Accuracy
+/- 1° (from 10 m
n
/sec and up) (At zero flow conditions, temperature
reading increases due to self-heating by the flow sensor)
Display
Technology
Liquid crystal
Back light
Blue with auto power save
Memory
2 Million point memory option
Mechanical
VPS.R080.M050
VPS.R250.M100
VPS.R01K.M200
135mm x 50mm x 85mm 0.7 Kg
135mm x 50mm x 85mm 0.7 Kg
155mm x 90mm x 125mm 1.6 Kg
IP grade
IP65 when mated to connector, at room temperature; direct rain and
sunlight should be avoided. Extreme temperature fluctuations may
affect the IP grade over time.
Ambient temperature
0..60°C
32 .. 140°F
Inputs and outputs
Analog
4..20mA or pulse, selectable via installation software
Serial IO
Modbus RTU
USB for configuration (display version only)
Supply
12..24 VDC +-10% CLASS 2 (UL)
Power consumption
150mA at 24VDC