© 2017 Van Putten Instruments BV | MAN-VP-SINL-UK | Revision:1701 | Date:22-03-2017
11
5.4
Totalizer
The totalizer keeps track of the total consumed amount of compressed air in normal cubic meters, or
in scf depending on which unit you choose to read out. The refresh interval is 1 second, actual
measurement data will be available on the display and by Modbus. For back up reasons, the totalizer
value is written to it's internal memory with an interval of 15 minutes. A power down might result in
maximum 15 minutes of totalizer data loss.
In bi-directional operation, negative flow is subtracted from the totalizer. The totalizer will count
backwards as the compressed air is delivered back to the supply side. The totalizer can only be
reset to zero. It's not possible to set it to a different value.
The display will show totalizer values up to 999.999,9 and will then become 0,0 independent of the
taken unit. This will not clear the internal totalizer.