Load cell connection 

The load cell single or double strain gauge bridge outputs shall be connected using 
shielded cable. It shall than be connected to the measuring equipment bridge mV/V 
input signal channel 
The bridge power shall be connected to E+ and E- outputs and the measuring equipment signal 
input shall be connected to the S+ and S- outputs.  
The cable should be routed at least 100 mm from other cables, so that electromagnetic 
interference is avoided. Cable shield is not connected to the load cell body and shall be 
grounded in the other end. The load cell connector housing is connected to the load cell 
body and the cable shield shall not be connected in the cable connector but be grounded 
in the other end. Cable shield is then grounded in one point only. 
For installation in an explosive gas/dust or mining area, only trained personnel may 
perform dimensioning of cables and barriers. A descriptive system document should be 
prepared by the system designer. 

 

* Deviations may occur in customer specific types. 
 

 
 

Connector pin-out and wires color code:

Electrical connection 

Connector type:

 (M12 or equivalent IP67 qualified) 

Cable:

 Shielded 4 or 8-wire 0,25mm

2

 cable through  IP67 qualified cable gland 

Connector pin number * 

Pin 1

: E+ (positive excitation).  Bridge 1 

Pin 3

: S+ (positive signal). Bridge 1 

Pin 2

: S- (negative signal). Bridge 1 

Pin 4

: E- (negative excitation). Bridge 1 

Pin 5

: E+ (positive excitation).  Bridge 2  

Pin 7

: S+ (positive signal). Bridge 2 

Pin 6

: S- (negative signal). Bridge 2 

Pin 8

: E- (negative excitation). Bridge 2  

Cable: Cable wire colour * 

Red

: E+ (positive excitation).  Bridge 1 

Green

: S+ (positive signal). Bridge 1 

White

: S- (negative signal). Bridge 1 

Grey

: E- (negative excitation). Bridge 1 

Brown

: E+ (positive excitation).  Bridge 2  

Blue

: S+ (positive signal). Bridge 2 

Yellow

: S- (negative signal). Bridge 2 

Pink

: E- (negative excitation). Bridge 2 

Содержание BLH NOBEL

Страница 1: ...Single channel Load cell KOSD KIMD KISD Double channel Load cell KOSD D KIMD D User manual...

Страница 2: ......

Страница 3: ...4 Load cell ATEX Label 4 Functional safety 5 Electrical connection and supply voltage 6 Load cell connection 7 Application examples 7 Mechanical installation and maintenance 8 Mechanical data 8 Append...

Страница 4: ...trial systems Its basic function is force measuring or weighing applications The strain gauge bridge output wiring is connected to an overall measurement system Changes to current manual version Detai...

Страница 5: ...it individually The two separate electrical circuits are insulated from each other The following KxxD D load cell configurations are available Single bridge load cell with connector or cable connectio...

Страница 6: ...wo 4 pin connectors or dual cable connections KxxD D with one 4 pin connector and single cable connection These load cells are approved for use in an explosive hazardous area provided that suitable in...

Страница 7: ...5X Issue 2 see appendix 2 Functional safety T V 968 FSP 1462 00 17 see appendix 2 Environmental conditions PARAMETER Min Typ Max UNIT Environmental protection IP rating IP67 Operating Temperature Tamb...

Страница 8: ...cific model They can be ordered either with a cable connector or with an integrated cable The safety description is labelled on the load cell For the D version the safety description and connection is...

Страница 9: ...48 year Maximum performance level PL d The achievable performance level assumes that the double channel load cell shall be connected to an overall measuring instrument in a category 3 structure with a...

Страница 10: ...ve gas dust or mining area only trained personnel may perform dimensioning of cables and barriers A descriptive system document should be prepared by the system designer Deviations may occur in custom...

Страница 11: ...cell KxxD D used in hazardous area is shown below The cable shield is not connected to the load cell body and shall be connected in the other end Connection to barrier or isolating IS unit is shown i...

Страница 12: ...revent the cylindrical load cell body from rotating in the supports Standardized adapters for some load cell types are available others can be custom designed and produced by Vishay Nobel On request t...

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ...IECEx web site http iecex iec ch Certificate number IECEx BAS 14 0015X Issue No 2 Function safety Certificate The functional safety certificate for the KxxD D Load cell can be found on T V Rheinland...

Страница 18: ...Box 423 SE 691 27 Karlskoga Sweden Phone 46 586 63000 Fax 46 586 63099 blhnobel se vpgsensors com www blhnobel com Document no 35218 Publication 600659R7 Vishay Nobel AB 2020 02 17 Subject to changes...

Отзывы: