Product overview
9
© 2016 Van Putten Instruments BV | MAN-VP-MPRO-UK | Revision:1602 | Date:20-10-2016
3.3
Safety system
Installation under pressure can be dangerous. Make sure that you understand the
safety system before installing the VPFlowScope M.
The safety system is specially designed for the VPFlowScope M. When the VPSensorCartridge is
in the compression fitting and the safety cable is attached with the locking ring fully tightened, the
probe can never exit the compression fitting due to the matching length of the cable. When the
VPSensorCartridge is installed on the right depth, the safety cable can be strained and locked. The
complete safety system is tested and certified for safety. Instructions on how to assemble the
safety system can be found in
chapter 6 mechanical installation
.
1.
Compression fitting:
0.5 inch with Teflon ferrule
2.
Safety plate:
Used to connect fitting and safety
system
3.
Auto brake:
The safety chain can be slide through
the auto brake system. Moving upwards is only
possible when the safety lock is pushed downwards
4.
Safety lock:
This nut can be used to lock or unlock
the safety release. Push down the lock to release
the safety chain
5.
Safety chain:
The safety chain can be attached to
the VPSensorCartridge
3.4
Configuration
The VPFlowScope M needs only one step to be ready for basic operation. It needs to know the
exact inner pipe diameter for accurate measurement,
wrong inner diameter will lead to very
significant misreadings
. The pipe diameter can be programmed with the keypad on the display
module, via the built-in Web interface or with the VPStudio configuration software. This software
suite is also used for configuration of the outputs and data logger. This software can be
downloaded from our website.
www.vpinstruments.com/downloads
. The exact inner diameter can
be measured using a wall thickness gauge, ask us for availability.
The default units are set to SI. For read out in Imperial units, use the keypad to change the units via
the menu. Go to menu->settings->display->rows.