Flow sensor
Measuring principle
Thermabridge™ Thermal Mass Flow sensor
Flow range
0(0.5) ... 150 m
n
/sec | 0 ... 500 sfps
Bi-directional flow
Optional
Accuracy
2% of reading under calibration conditions
Reference conditions
0 °C, 1013.25 mbar | 32 °F, 14.65 psi
Gases
Compressed air, Nitrogen and inert, non condensing gases
Gas temperature range
0 ... 60 °C | 0 ... 140 °F
Pressure sensor
Pressure range
0 ... 10 bar | 0 ... 145 psi gage
Accuracy
+/- 1% FSS (total error band)
Temperature compensated
Temperature sensor
Temperature range
0...60 °C | 32...140 °F
Accuracy
> 10m/sec: +/- 1 °C | 1.8 °F
< 10m/sec: + 5 °C | 9 °F
Mechanical & Environmental
Probe Length
340 mm | 13.4"
Weight
200 grams | 7.05 ounces
Process connection
Compression fitting, 1/2" NPT, Tapered
Pressure rating
PN10
Protection grade
IP65 | NEMA 4 when mated to transmitter
Ambient temperature range
0...60 °C | 32...140 °F, avoid direct sunlight or radiant heat
Wetted materials
Anodized Aluminum, Stainless steel 316, Glass, Epoxy
Corrosion resistance
Highly corrosive or acid environments should be avoided
Electrical
Connection type
VPSensorCartridge proprietary
Power consumption
See transmitter specifications for combined power consumption