98-100 ISSUE 8
27
8. Enter the terminal names between which the inductance will be measured.
By pulling down the menu along side the fill-in box, you may select the
high and low terminals for each test.
9. Select (mouse click the button) the type of limits you require, and enter the
values.
%
Enter a nominal value and then the limits as negative and
positive values.
><
Enter minimum and maximum values.
>
Enter just a minimum value.
<
Enter just a maximum value.
10. Select OK. The test specification will then be displayed in the program
window.
ZBX – I
MPEDANCE WITH
E
XTERNAL
B
IAS
The Winding Impedance with External Bias test measures the impedance of a
selected winding while applying a DC current from the DC1000 through the
winding. An AC voltage is also applied across the winding from the AT.
This test can be used with inductors to measure the change in impedance with
a bias current.
On selecting “ZBX Impedance with External Bias” from the ‘available tests’
window, the following dialog box is displayed: