background image

KBM-110

Safety | 3

B-H86.0.12.DB214-1.0

7 / 32

3 Safety

3.1 Explanation of safety symbols

 DANGER

This symbol warns of imminent danger which can result in death, severe bodily injury,
or severe property damage in case of non-observance.

 DANGER

This symbol indicates danger for living tissue as well as a variety of materials, which
can be damaged or destroyed when coming into contact with this chemical. Caustic
effect, protective equipment required!

 DANGER

This symbol indicates danger for all life forms, which can result in death or acute or
chromic damage to the health after inhaling, swallowing or absorbing this chemical
through the skin.

 CAUTION

This symbol warns of potential dangers or harmful situations which can cause damage
to the device or to the environment in case of non-observance.

 NOTE

This symbol indicates processes which can have a direct influence on operation or
can trigger an unforeseen reaction in case of non-observance.

 NOTE

This symbol instructs the use of eye protection which protects the eyes from harmful
influences when working with powerful light, UV radiation, laser, chemicals, dust,
splinters or weather influences.

 NOTE

This symbol instructs the use of protective gloves which offer protection from mechan-
ical, thermal, chemical, biological or electrical hazards.

3.2 Foreseeable misuse

The fault-free function and operational safety of the product can only be guaranteed if
generally applicable safety precautions and the device-specific safety instructions for
this document are observed.

If these notices are disregarded, personal injury or death, as well as property damage
can occur.

Содержание KBM-110

Страница 1: ...Operating manual EN KBM 110 pH Tester BNC connection Waterproof Precise and fast...

Страница 2: ...5 The product at a glance 11 5 1 The KBM 110 11 5 2 Display elements 11 5 3 Operating elements 11 5 4 Connections 12 6 Bases for measurement 13 6 1 pH measurement 13 6 1 1 Explanation 13 6 1 2 pH elec...

Страница 3: ...tion menu 22 8 2 3 Configuring parameters of the configuration menu 23 8 2 4 Adjustment of the measuring input 25 8 2 5 Configuring parameters of the adjustment menu 26 9 Error and system messages 28...

Страница 4: ...egal address of the manufacturer KBM 110 4 32 B H86 0 12 DB214 1 0 1 Legal address of the manufacturer Conrad Electronic SE Klaus Conrad Str 1 D 92240 Hirschau http www conrad com WEEE reg no DE 28001...

Страница 5: ...king safely and efficiently with the product In addition to the quick reference guide with all relevant legal and safety content in hard copy this document is a detailed reference option for the produ...

Страница 6: ...nical controls Product functions Product labels Cross reference p 5 Foot notes 2 5 Further information Software version of the product V1 2 or later For the exact product name refer to the type plate...

Страница 7: ...which can cause damage to the device or to the environment in case of non observance NOTE This symbol indicates processes which can have a direct influence on operation or can trigger an unforeseen r...

Страница 8: ...foods Inspect the electrode before and after the measurement Always measures in samples for measurements in foods Discard these samples after the measurement DANGER Potassium chloride potassium nitra...

Страница 9: ...plication examples for this are for example drinking water waste water surface water swimming pools fish breeding and process chemistry See Technical data p 31 3 5 Qualified personnel For commissionin...

Страница 10: ...d reliability in a compact ergonomic housing Additional impressive features include the dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which offers ov...

Страница 11: ...r value for min max hold Auxiliary display Corresponding temperature for the displayed pH value with unit Measured temperatures are dis played with a decimal place adjusted without Bar graph Progress...

Страница 12: ...ion ONF appears in the dis play Long press 4s Start automatic calibration AL appears in the dis play 5 4 Connections BNC connection Connection for electrode Un locking with rotating ring on the cable...

Страница 13: ...roof versions from exten ded immersion above the shaft calibrate electrodes sufficiently often The can range from every hour to several weeks depending on the electrode and the application Use a suita...

Страница 14: ...electrode with two ceramic diaphragms and liquid electrolyte GE 104 BNC is preferably used for measurements in low ionic media such as rainwater aquarium water and deionised water GE 114 WD is a unive...

Страница 15: ...he pH probe must always be protected Plastic shafts must not be cleaned in solvents etc If in doubt contact the manufacturer to inquire about suitable cleaners for the existing electrode This is also...

Страница 16: ...ld always be stored wet in 3 mol l KCl Extended storage in distilled or deionised water will result in depletion of the reference electrolytes NOTE The pH electrode included in the scope of supply sho...

Страница 17: ...ce the batteries Prerequisites The product is switched off A suitable PH1 is available Instruction 1 Unscrews the Phillips screws and remove the cover 2 Carefully replace the two Mignon AA batteries E...

Страница 18: ...ible in a temperature range from 0 C to 60 C We recom mend performing calibration at temperatures between 10 C and 40 C NOTE Calibration should be conducted at the same temperature used for the measur...

Страница 19: ...tomatically Prerequisite The product is switched on The pH electrode and if applicable a temperature sensor are inserted in the product Ready to use GPH 7 0 buffer solution Ready to use GPH 4 0 or GPH...

Страница 20: ...uation is not possible in this case We recommend conducting a 2 point calibration because a 1 point calibra tion only entails a shift of the zero point Prerequisite An arbitrary buffer solution is ava...

Страница 21: ...second buffer solution with a value below pH 6 or above pH 8 NOTE A gradient compensation with buffer solutions between pH 6 and pH 8 is not possible With entry of the compensation value the resultin...

Страница 22: ...n of the temperature sensor was made AL Calibration Blinks if no valid calibration is available The product is now ready for measurement NOTE The product must be calibrated to the electrode prior to s...

Страница 23: ...oduct is switched off without saving the configuration the last save value is re produced on the next start up of the product 8 2 3 Configuring parameters of the configuration menu Description The fol...

Страница 24: ...ure display in F Alarms AL OFF No active alarm ON Alarm alerting via text display acoustic signal and flashing of the backlighting BEEP Alarm alerting via text display and acoustic signal L TE Alarm a...

Страница 25: ...tment is made you change the pre adjusted factory settings This is signalled with the T OF or T SL when the product is switched on The standard settings of the zero point value and the gradient value...

Страница 26: ...he following representation Representation Parameter Values Meaning Zero point correction T OF 0 00 No zero point correction 5 00 5 00 Zero point correction in C and or at F 9 00 9 00 Gradient correct...

Страница 27: ...KBM 110 Operation 8 B H86 0 12 DB214 1 0 27 32 NOTE If the product is switched off without saving the configuration the last save value is re produced on the next start up of the product...

Страница 28: ...ontamin ated or defective Use fresh buffer solution Clean electrode re calibrate Replace electrode AL ERR 3 Slope is too high Incorrect buffer solu tion used Buffer solution is contaminated Electrode...

Страница 29: ...KBM 110 Error and system messages 9 B H86 0 12 DB214 1 0 29 32 Replace batteries Send in for repair...

Страница 30: ...se take it to a municipal collection point where it will be transported to a dis posal company in accordance with requirements of hazardous goods laws Otherwise return it to us freight prepaid We will...

Страница 31: ...d weight 108 54 28 mm without BNC plug 130 g incl battery without electrode 190 g incl battery and electrode Operating conditions 20 to 50 C 0 to 95 r h temporarily 100 r h Storage temperature 20 to 7...

Страница 32: ...Manufacturer If you have any questions please do not hesitate to contact us Contact VOLTCRAFT Distributed by Conrad Electronic SE Klaus Conrad Str 1 92240 Hirschau Germany Tel 49 9604 40 87 87 Fax 49...

Отзывы: