d) Advanced presets
The meter provides some presets to facilitate and improve operation in different environments.
The following presets can be made: The bold values in the table are factory presets.
Parameters Value
Meaning
PoFF
yES
no
Automatic shut-off setting
yES = on / no = off
bEEP
yES
no
Beep setting
yES = on / no = off
t-CF
C
F
Temperature unit setting
C = °C / F = °F
SALt
0 - 50
Adjustment of salt content (salt compensation) only relevant
when measuring dissolved oxygen 0 – 50 = 0% to 50%,
setting step 1%
High
0 - 8900
Adjustment of height compensation in meters only relevant
when measuring dissolved oxygen 0 – 8900 = 0 m – 8900 m,
setting step 100 m
Highf
0 - 29300 Adjustment of height compensation in feet (ft) only relevant
when measuring dissolved oxygen 0 – 29300 = 0 ft – 29300 ft,
setting step 100 ft
ESC
Exits the presets
To open the settings menu, press and hold down the “SET” button (6) for approx. 2 seconds.
The menu will open.
Use the “SET” button (6) to switch parameters.
Use the “▲” (5) and “▼” (8) buttons to set values.
Each parameter setting must be confirmed with the “ENTER” button (7). After that, the preset
parameters are retained even after turning off/battery change.
To exit the settings menu, press the “ESC” button (3) or select the “ESC” parameter and press
the “SET” button. The settings menu will close and the measurement display will appear.