41
V. Measuring
5.8.3
Volume
Here you can program, how much titrant is to be
injected per dosing step.
Fixed:
The pump will deliver the same amount of titrant in
every dosing step.
Dynamic:
Selecting this mode, the dosage will be set under dyna-
mic control. Parameters for controlling the dosage are
ΔE, Vmin and Vmax. ΔE is the mV signal change bet-
ween two dosage steps, and Vmin Vmax specify, how
the titrant pump should respond to this signal change.
Example: ΔE be set to 8 mV. If the signal changes more
than 8 mV, the pump will deliver less titrant with a mini-
mum of Vmin. If the signal change is smaller than 8 mV,
the pump will deliver more titrant with a maximum of
Vmax.
Hint
If you want titrations to be more precise, go for lower
values in ΔE, Vmin and Vmax. And vice versa, if not
accuracy but speed is the primary goal: set ΔE, Vmin
and Vmax to higher values.
Содержание BTG Mutek PCD-06
Страница 1: ...Operation Manual M tekTM PCD 06 Light Standard Premium Particle Charge Detector...
Страница 6: ...List of contents Chapter Paragraph Title Page 6 Operation Manual M tekTM PCD 06 BTG...
Страница 20: ...20 Operation Manual M tekTM PCD 06 BTG 2021 II Product description...
Страница 24: ...22 Operation Manual M tekTM PCD 06 BTG 2021 III System layout and functions...
Страница 32: ...30 Operation Manual M tekTM PCD 06 BTG 2021 V Measuring 5 3 Overview Main Screens...
Страница 56: ...54 Operation Manual M tekTM PCD 06 BTG 2021 VIII Shutting down the system storage and disposal...
Страница 59: ......
Страница 60: ...BTG Instruments AB P O Box 602 SE 661 29 S ffle Sweden Phone 46 533 426 00 www btg com...