V. Measuring
5.8.1
Setting of standard methods
Method 900:
Method 901:
Editable method parameters are as follows:
Termination
mV
EP
0 mV
Pump
Automatic
Type
EP
Meas. Mode
fixed
Increment
3 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,01 ml
dV (max)
0,1 ml
Termination
pH
EP
7,00 pH
Pump
Automatic
Type
IEP
Meas. Mode
Adaptive
dE
1,0 mV
dt
1 s
t (min)
3 s
t (max)
15 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,005 ml
dV (max)
0,05 ml