© BTG 2021
Operation Manual Mütek
TM
PCD-06
21
III. System layout and functions
3.1.1
Measuring cell
Standard delivery of the Mütek™ PCD-06 includes a
precision measuring cell and a displacement piston. Fit-
ted in the measuring cell are two platinum electrodes for
identifying the streaming potential.
Fig. 3-3
Measuring cell and displacement piston
3.1.2
pH set (optional)
The Mütek™ PCD-06 Smart identifies not only the
charge level but also the pH of a sample.
In combination with a Travel Titrator module or an exter-
nal Titrator, the instrument can determine the isoelectric
point of a sample via the pH electrode.
Install the pH electrode in the measuring cell on the wall
opposite the titrant tubes, that is at a maximum distance
from the tube tips.
Fig. 3-4
Measuring cell and pH electrode
Designations
1
Electrode cable with A2 connector
2
Plexiglass shaft
3
Clip
4
Ceramic diaphragm
5
Electrode cap
The pH electrode must immerse in the sample solution.
Therefore it is necessary to use an increased sample
volume of 30 ml for isoelectric point measurements. Use
more sample material and do not dilute the sample with
water!
When just one pH value has to be measured (no pH
curve or IEP measurement) the pH of the 10 ml sample
can be measured outside before filling into the measu-
ring cell.
1
2
3
4
5
Содержание BTG Mutek PCD-06
Страница 1: ...Operation Manual M tekTM PCD 06 Light Standard Premium Particle Charge Detector...
Страница 6: ...List of contents Chapter Paragraph Title Page 6 Operation Manual M tekTM PCD 06 BTG...
Страница 20: ...20 Operation Manual M tekTM PCD 06 BTG 2021 II Product description...
Страница 24: ...22 Operation Manual M tekTM PCD 06 BTG 2021 III System layout and functions...
Страница 32: ...30 Operation Manual M tekTM PCD 06 BTG 2021 V Measuring 5 3 Overview Main Screens...
Страница 56: ...54 Operation Manual M tekTM PCD 06 BTG 2021 VIII Shutting down the system storage and disposal...
Страница 59: ......
Страница 60: ...BTG Instruments AB P O Box 602 SE 661 29 S ffle Sweden Phone 46 533 426 00 www btg com...