![VNS Therapy SENTIVA M2000 Wand Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/vns-therapy/sentiva-m2000-wand/sentiva-m2000-wand_technical-manual_1043386060.webp)
60
Permissable Scan Area
C7
L3
MRI Exclusion Zone
Scan Conditions - SenTiva® Model 1000
No special MRI equipment/coils required
GROUP A
Generator in upper
left chest, at or above
armpit (above rib 4)†
Note:
The scan iso-center must be
outside the exclusion zone
Imaging techniques such as computed tomography, x-ray, and
ultrasound are safe to perform in the MRI exclusion zone.
Review the most current labeling prior to performing an MRI scan.
For full MRI safety information, refer to MRI Instructions for Use at
www.easy-mri.com
† Patients with implants in other locations
must follow Group B scan conditions
SenTiva MRI Guidelines