52
Turn Tachycardia Detection ON
Tachycardia Detection can be turned on from the Edit Parameters screen.
To turn Tachycardia Detection ON:
1.
From the Summary Screen, tap on the
EDIT PARAMETERS
button,
OR tap on the
PARAMETERS
button in the Navigation Bar at the bottom
of the screen
2.
On the Parameters Screen, select the Detection Tab and then tap on
ENABLE NOW
to turn on Tachycardia Detection
Perform System Diagnostic #1
System Diagnostic #1 tests the continuity of the lead and confi rms delivery of
the programmed output current to the vagus nerve.
To perform System Diagnostic #1:
1.
From the Summary Screen, tap on the
PERFORM DIAGNOSTICS
button,
OR tap on the
DIAGNOSTICS
button in the Navigation Bar
2.
On the Diagnostics screen, tap
TEST SYSTEM
to start the test and confi rm
that the system is working correctly
3.
Review the Diagnostics Test results and tap
DONE