05_2018
25
Deleting the saved values: CLR
(blocked)
In this menu item, you can reset the
values in a memory location to the
factory default settings.
Factory settings:
•
OFSET = 0 %
•
GAIN = 1.0
•
LIM 1 = 18,5 %
•
LIM 2 = 24 %
This menu option has no settings. It is
simply activated, setting back the set
-
tings. Activate it using the MENU-but
-
ton.
Easy Check: ECH (blocked)
Easy Check is a quick functional check
of the device. It does not replace a cali
-
bration, but can be performed anywhe
-
re, any time.
HINT: to perform the EasyCheck, the
reference oil CO 330 is required. It can
be purchased separately.
Please follow the instructions of the
reference oil CO 330 to perform the
Easy-Check.
Ther result is „OK“ for passed or „NOK“
for not passed.
”NOK„
English