Instructions for Use
75
8
Troubleshooting
For technical enquiries, please contact your distributor or Vitl at the address
given on page 2 of this manual.
Table 24: Troubleshooting Suggestions
Unit will not turn on
•
No power
→
Check power is switched on at wall outlet socket
•
Bad USB connection
→
Ensure power connector fully inserted into rear of unit
Unit spuriously resets or restarts automatically
•
Power connecter loose
→
Ensure USB connecter fully inserted at rear of unit
•
Bad USB cable
→
Use good quality USB 2.0 cable, as supplied with the unit
•
Supply brown-outs or black-outs
→
Power unit from a stable AC mains supply
Unit does not recognise sample tube insertion
•
Tube too small →
Ensure minimum tube requirements met
4.1
Unit Protocol and Timer buttons are greyed-out
•
Clock not set
→
Ensure the Date and Time are set
Measurement result lower than expected
•
Tube poorly inserted
→ Ensure tube is pushed fully down in tube holder
4.1
•
Low sample volume →
Ensure sample greater than minimum operating volume
4.1
•
Wrong protocol selected
→ Ensure using correct measurement protocol
6.4
•
Dirty chamber
→ Check that dirt not built up in sample measurement chamber
9.1
Measurement result higher than expected
•
Wrong protocol selected
→ Ensure using correct measurement protocol
6.4
•
Light ingress →
Do not use unit in direct sunshine or under infrared lighting
Lu-mini App reports No Units Attached
•
USB Driver
→ Ensure
that the USB Driver is installed