The users of the device must not be under the influence of alcohol or drugs.

 

 

 

The  predicted  time  of  using  the  device  is  4  years.  After  this  period  it  can  be  used  only  after  a  positive  control 
performed by the manufacturer.

 

 

The device may cause decolouration of the floor due to rubber elements during normal exploitation.

 

 

2.8

 

Environmental protection 

 

The device excluding the upholstery is made of steel and undergoes recycling In 95%. 

 

After delivery the service is to collect the packaging. All elements made of polythene may be totally recycled. 

 
 

2.9

 

Periodical servicing 

 

Table 2 constitutes a periodical servicing card, which includes inspections of the device that need to be performed 

periodically. Inspections and technical maintenance should be performed according to the schedule specified in the card. 

 

Table 2. 

INTERVAL 

ELEMENT 

INSPECTION 

METHOD OF REPAIR 

 

 

Check visually the state of 

 

 

 

parapodium, if there are any visible  If any damages are 

 

Dynamic 

damages, such as fissures, 

found, replace 

 

Parapodium 

deformations or any other visible 

damaged elements with 

 

 

damages, which can affect device 

new ones. 

 

 

usage or safety of the user. 

 

 

 

Measure height of rubber veneer 

 

 

 

Minimum permissible height – 5 

 

 

 

mm 

 

 

Internal feet ** 

 

Replace 

 

PD0325 4 pieces 

 

 

 

 

 

 

 

 

6 months * 

 

Measure height of rubber veneer 

 

 

 

Minimum permissible height – 

 

 

 

2mm 

 

 

External feet ** 

 

 

 

PD0326 2 pieces 

 

Replace 

 

PD0327 2 pieces 

 

 

 

 

 

 

 

 

Check if all regulation screws, nuts  Follow "Assembly and 

 

 

and bolts are fastened 

Adjustment Instruction" 

 

ATTENTION 

 

 

 

 

 

 

 

Lower 

If any fissures, cuts or permanent 

 

 

articulation shock  deformations of lower articulation 

Replace 

 

absorber 

shock absorbers are found – 

 

 

12 months * 

PD0442 4 pieces  replace 

 

 

Corset’s shock 

If any fissures, cuts or permanent 

 

 

absorber 

deformations of corset’s shock 

Replace 

 

PD0361 4 pieces  absorbers are found – replace 

 

 

*

 

time of intense usage - 2h/day 

 

**

 

if used on hard and rough surfaces (e.g. concrete floor) check every 2 months 

 

In case of the malfunction of the parapodium it should not be used and the service should immediately be informed. 
The name and the address of the authorised service is on the Warranty Chart. 

 
 
 

 

14 

Содержание PARAPODIUM PD 180

Страница 1: ...Dynamic Parapodium Model PD150 PD 180 PD 200 User s Manual 09 2016...

Страница 2: ...PRELIMINARY REHABILITATION PROGRAMME FOR THE PERIOD OF ADAPTATION OF THE ORGANISM TO THE DEVICE 5 2 USER S MANUAL 6 2 1 GENERAL PRINCIPLES OF SAFETY 6 2 2 GENERAL INFORMATION 7 2 3 PREPARATION OF THE...

Страница 3: ...ralysis allowing to assume upright position without additional support on crutches and sitting up Dynamic Parapodium parapodium orthosis system making active rehabilitation and movement gait possible...

Страница 4: ...4 Most Common Systemic Disturbances Likely To Occur In Paraplegia WARNING The prerequisite for starting rehabilitation of the patient making use of the Dynamic Parapodium is previous consultation wit...

Страница 5: ...n and inappropriate distribution of blood which leads in high temperature of the environment to the earlier occurrence of the over warming effect than in healthy subjects associated with the risk of d...

Страница 6: ...round During the brakes in the previous exercise inclining and turning the trunk in various directions and or active training of the upper extremities with various loads 3 Prolongation of the period o...

Страница 7: ...situations or serious injury to the user 2 2 General information The purpose of this manual The manual contains detailed information concerning the conditions of usage of the Dynamic Parapodium Model...

Страница 8: ...lanes The plane perpendicular to the axis of the runners extending through the corset hanger joint after positioning the patient in the parapodium this plane extends through the axis of their hip join...

Страница 9: ...unners from the ground is sufficient to initiate the movement of the parapodium The runners are equipped with lateral safety supports It is very difficult to make parapodium swing beyond the support p...

Страница 10: ...ient feels safe support for the leg but without hyperextension of knee joints This ensures not only safety but also correct position of the body in the parapodium which makes walking much easier The p...

Страница 11: ...relation to the axis of corset hanger joints if the patient is bent forward act according to Assembly Instructions if the patient is bent backward act according to Assembly Instructions WARNING Durin...

Страница 12: ...the parapodium is assembled incorrectly act according to Assembly Instruction When the patient moves the trunk and lifts the runners from the ground the front of the runners fall down Assembly Instruc...

Страница 13: ...ld appliances is acceptable Leather upholstery should be cleaned and conserved with special agents used to clean leather clothing Velour upholstery should be washed at 40 C with water containing mild...

Страница 14: ...e Dynamic damages such as fissures found replace Parapodium deformations or any other visible damaged elements with damages which can affect device new ones usage or safety of the user Measure height...

Страница 15: ...ease share your opinion with us and we will consider introducing the appropriate changes It is a great challenge for mdh sp z o o to create a perfect dynamic orthosis that could serve all disabled peo...

Страница 16: ...24 www viteacare com Dynamic Parapodium Model PD150 PD180 PD200 User s Manual is a property of mdh Sp z o o All rights reserved Attention The devices presented in this manual are protected by appropri...

Отзывы: