ENG
32
Page
of 36
DT_0006 Issue 15
A Global Leader in Respiratory Solutions
EN 60601-1-2:2015 - Emissions tests
Emissions
test
Compliance
Electromagnetic environment -
guidance
RF
emissions
CISPR 11
Group 1
The VitaloJAK uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
RF
emissions
CISPR 11
Class B
The VitaloJAK is suitable for use
in all establishments, including
domestic establishments and
those connected to the public
mains network (e.g. at home and
doctor’s offices in residential areas)
EN 60601-1-2:2015 - Immunity tests
Immunity test
Test level
Compliance level
Reached
Electrostatic
discharge (ESD)
EN 61000-4-2
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Contact: ± 8 kV
Air: ± 2 kV,± 4 kV,
± 8kV,± 15 kV
Power frequency
(60Hz) magnetic
field IEC 61000-4-8
30A/m
30A/m
Radiated RF
EN 61000-4-3
10 V/m
80MHz to
2700MHz
10 V/m
80MHz to 2700 MHz
Proximity fields
from RF devices
EN 61000-4-3
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
9 to 28V/m
385 to 5785MHz
As per Table 9
EN60601-1-2:2015
Electrostatic Discharge Warning:
Although the VitaloJAK is
designed to be unaffected by typical electrostatic discharge (ESD)
when in recording mode and in its pouch, when the device is not
in its pouch very high levels of ESD can result in a temporary
suspension of normal operation, requiring the operator to turn
Содержание VitaloJAK 7100
Страница 36: ......