Incorrect placement – incisal (fig. 17):
Probe tip must be a minimum of 2 mm
from the incisal edge.
Incorrect placement – cervical (fig. 18):
Probe tip too close to gingival tissue.
Maintain a minimum distance of 2 mm to the tissue margin.
If the probe tip is moved off of the tooth prior to the instrument
sounding a “beep”, an error message or an erroneous measurement
will be displayed and the measurement will need to be repeated.
Note:
Prior to measuring a tooth, an anti-infection cover must be
applied to the VITA Easyshade Compact probe and the instrument must
be calibrated. It is recommended to remove any stains from the surface
of a tooth prior to measurement. It may be required to wipe the tooth
with gauze to prevent the tip from sliding. The probe tip must be placed
entirely on the natural tooth. Any contact with existing restorations
must be avoided. Underlying restorations may adversely affect the
measurement result.
Measuring a tooth