VistaVision 82026-630 Скачать руководство пользователя страница 7

Operating Manual of Inverted Microscope 

 

 

- 3 - 

There are 2 styrofoam boxes within the carton box.   

a)

 

The  first  box  contains  6  objectives  (3  pcs  plan  objectives  10X,  25X  &  40X  with  single 
colored ring and 3 pcs plan phase contrast 10X, 25X & 40X with double colored rings and 
marked 

PHP2

),  1  trinocular,  2  eyepieces,  1  centering  telescope,  3  filters  ,  1  spare  fuse 

and1 spare lamp. 

b)

 

The  Second  box  contains  1  body  frame,  1  phase  contrast  condenser,  3  different  size  stage 
brackets holders. 

 

     

Installation steps

 

The  number  in  the  following  installation  steps  is  shown  in  the  installing  diagram,  such  as 

1

  is 

Body

 in Fig.2 

(1) Unpack this unit, take out the body frame 1 and place it on a stable work bench, and remove the 

supporting package and the dust cap (bag). 

(2)  Take  out  the  phase  contrast  condenser  4.  Loosen  the  screw 

  on  the  condenser  holder  3  and 

then install the condenser into the holder, tighten the screw again. As shown in Fig. 3. 

(3) Take out the stage bracket 2 (when the stage bracket is used to carry a specimen or observe tissue 

in the culture dish), put the desired specification stage bracket flatly in the stage  moving device 
rack 

. As shown in Fig. 3. 

 
 
 
 
 
 
 
 
(4) Remove the dust cap of eyepiece tube, insert the eyepiece 5 into the eyepiece tube, and turn the 

eyepieces so that it fits the eyepiece tube well. 

(5)  There  are  6  objectives.  Those  with  single  colored  ring  are  plan  objectives  and  with  double 

colored rings are phase contrast objectives which are also marked 

PHP2

. Install them onto the 

nose piece.  User  can  only  mount  5 objectives onto the nose  piece  at  any  one time. Install  them 
clockwise in the ascending order of magnification As shown in Fig.4. 

 
 
 
 
 
 
 
 

(6) Connect the power cord to the power supply socket of the body.   
(7) Check if the above installation is secure and safe. 
(8)  Inspect  and  gather  the  accessories  and  tools  enclosed  in  the  package  and  keep  them  in  a  safe 

place to avoid misplacement. 

Fig.3 

 

 

Fig.4 

Содержание 82026-630

Страница 1: ...Inverted Microscope Model 82026 630 Operating Manual...

Страница 2: ...Our product has been design to provide the highest level of safety However improper operation or negligence in following the instructions in this manual may cause personal injuries and property losses...

Страница 3: ...nded PE 4 Do not expose unit to high temperatures and dampness Suitable operating environment is between 5 C to 35 C and relative humidity of 20 80 at 25 C Do not subject unit to direct sunlight 5 Do...

Страница 4: ...e 3 Table of Contents I Characteristics and applications 1 II Structural features 1 III Installation of this unit 2 IV Technical specifications 4 V Operation 5 VI Replacement of bulb and fuse 13 VII M...

Страница 5: ...the light path to perform transmitted phase contrast microscopic observation This unit is suitable for the microscopic observation of cells tissues and clear liquid tissues and the dynamic microscopi...

Страница 6: ...Operating Manual of Inverted Microscope 2 III Installation Caution Before installing be sure every components is clean no score any parts or glass surface 1 2 2 2 4 6 5 8 3 7 Fig 2...

Страница 7: ...condenser into the holder tighten the screw again As shown in Fig 3 3 Take out the stage bracket 2 when the stage bracket is used to carry a specimen or observe tissue in the culture dish put the des...

Страница 8: ...78 1 2 Objectives 40X 0 60 3 32 1 2 10X 0 25 8 80 1 2 Remark PHP2 25X 0 40 4 78 1 2 Remark PHP2 Phase contrast objective 40X 0 60 3 32 1 2 Remark PHP2 Ultra long working distance condenser Working di...

Страница 9: ...tment of the condenser height Before this unit is released from the factory the condenser already bears a calibrated height mark Turn the condenser height control knob to move the condenser into worki...

Страница 10: ...enter distance of the eyepiece tubes by turning the left or right frame body until the field of view becomes a fully overlapped circle as shown in Fig 8 b 6 Inspection of transmitted lighting path The...

Страница 11: ...enser center center of aperture diaphragm must be adjusted The center of the light spot can coincide with that of the field of view by adjusting two the centering screws of the condenser holder as sho...

Страница 12: ...be removed as shown in Fig 13 This cannot be done until the bulb is cool Wrap the bulb with a cotton fabric and adjust the bulb vertically gently in the direction indicated by the arrow in Fig 13 This...

Страница 13: ...gh the eyepiece Turn the coarse focusing control knob slowly to lower the objective When a specimen image appears in the field of view stop turning the coarse focusing control knob as shown in Fig 16...

Страница 14: ...fting mechanism is lowered Use a screwdriver to loosen the 3 mounting screws at the bottom of the stage moving device remove the stage moving device and lay it flatly on the worktable upside down Do n...

Страница 15: ...phase contrast objective into the light path and turn the phase contrast turntable or phase contrast puller to the 10 scale Turn condenser height control knob so that the phase contrast condenser rise...

Страница 16: ...dge clear 4 In the phase contrast observation mode the center of the dark ring should coincide with that of the bright ring If their centers do not coincide as shown in Fig 20 c it is necessary to cen...

Страница 17: ...b and its surroundings have cool down This is to prevent hand getting burnt by heat 3 Unscrew the screw loosen the screw and pull out the lamp holder from the halogen lamp housing Use a cloth to remov...

Страница 18: ...nless the electric components in this unit is still operating When this unit is not to be used for a long time remove the power plug from the supply socket and keep all cables properly 2 This unit sho...

Страница 19: ...a conforming halogen lamp Optical system and imaging The nosepiece has not been turned to the fixed position Turn the nosepiece to the fixed position The filament image deviates from the center of th...

Страница 20: ...not laid flatly Lay the specimen flatly on the object stage and hold it stably The phase contrast objective does not match the ring plate of the phase contrast condenser lens Align the phase contrast...

Отзывы: