VistaVision 82026-630 Скачать руководство пользователя страница 10

Operating Manual of Inverted Microscope 

 

 

- 6 - 

3. Adjustment of phase-contrast device 

Set the phase contrast puller 

 to the central light hole, as shown in Fig.7. 

 

 

 

 

 

4. Reset of diopter adjustment ring 

Turn  the  d

iopter  adjustment

  ring 

  on  the  left    eyepiece  tube,  so  that  the 

0

  d

iopter

  position  is 

aligned with the side scale (remark white line), as shown in Fig.8. 

5. Adjustment of interpupillary distance 

Parallax  can  be  eliminated  by  adjusting  the  interpupillary  distance  so  that  the  distance  of  the 
eyepiece  tube  is  identical  with  your  interpupillary  distance  and  enable  you  to  observe  more 
comfortably and clearly. When you observe through two eyepieces, if the field of view consists of 
two  overlapping  circles,  as  shown  Fig.8-a,  you  can  alter  the  exit  pupil  center  distance  of  the 
eyepiece  tubes  by  turning  the  left  or right  frame  body 

  until  the  field  of  view  becomes  a  fully 

overlapped circle, as shown in Fig. 8-b. 

 

 

 

 

 

 

 

6. Inspection of transmitted lighting path 

The  optical  system  of  this  unit  has  been  calibrated  before  factory  release.  But  the  center  of  the 
lighting  bulb  or  condenser  system  may  deviate  from  the  center  of  the  optical  system  due  to  any 
possible  violent  vibration  or  inclination  during  transport.  After  the  above  installation  and 
adjustment, the light path of the optical system should be checked. 

(1) Inspection of center of field diaphragm 

A.  Put  the  test  slide 

.as  an  example  in  the  selected  Stage  bracket 

.  Turn  the  10X  plan 

objective 

  into the light path. Turn the stage moving device control knob and 

  and 

 

to move the slide right above the 10X objective. Adjust the focus by the coarse and fine focus 
knobs 

  and 

    to make the image of slide in the field of view clear. As shown Fig.9. 

 

 

Fig. 7 

 

 

Fig. 8 

Содержание 82026-630

Страница 1: ...Inverted Microscope Model 82026 630 Operating Manual...

Страница 2: ...Our product has been design to provide the highest level of safety However improper operation or negligence in following the instructions in this manual may cause personal injuries and property losses...

Страница 3: ...nded PE 4 Do not expose unit to high temperatures and dampness Suitable operating environment is between 5 C to 35 C and relative humidity of 20 80 at 25 C Do not subject unit to direct sunlight 5 Do...

Страница 4: ...e 3 Table of Contents I Characteristics and applications 1 II Structural features 1 III Installation of this unit 2 IV Technical specifications 4 V Operation 5 VI Replacement of bulb and fuse 13 VII M...

Страница 5: ...the light path to perform transmitted phase contrast microscopic observation This unit is suitable for the microscopic observation of cells tissues and clear liquid tissues and the dynamic microscopi...

Страница 6: ...Operating Manual of Inverted Microscope 2 III Installation Caution Before installing be sure every components is clean no score any parts or glass surface 1 2 2 2 4 6 5 8 3 7 Fig 2...

Страница 7: ...condenser into the holder tighten the screw again As shown in Fig 3 3 Take out the stage bracket 2 when the stage bracket is used to carry a specimen or observe tissue in the culture dish put the des...

Страница 8: ...78 1 2 Objectives 40X 0 60 3 32 1 2 10X 0 25 8 80 1 2 Remark PHP2 25X 0 40 4 78 1 2 Remark PHP2 Phase contrast objective 40X 0 60 3 32 1 2 Remark PHP2 Ultra long working distance condenser Working di...

Страница 9: ...tment of the condenser height Before this unit is released from the factory the condenser already bears a calibrated height mark Turn the condenser height control knob to move the condenser into worki...

Страница 10: ...enter distance of the eyepiece tubes by turning the left or right frame body until the field of view becomes a fully overlapped circle as shown in Fig 8 b 6 Inspection of transmitted lighting path The...

Страница 11: ...enser center center of aperture diaphragm must be adjusted The center of the light spot can coincide with that of the field of view by adjusting two the centering screws of the condenser holder as sho...

Страница 12: ...be removed as shown in Fig 13 This cannot be done until the bulb is cool Wrap the bulb with a cotton fabric and adjust the bulb vertically gently in the direction indicated by the arrow in Fig 13 This...

Страница 13: ...gh the eyepiece Turn the coarse focusing control knob slowly to lower the objective When a specimen image appears in the field of view stop turning the coarse focusing control knob as shown in Fig 16...

Страница 14: ...fting mechanism is lowered Use a screwdriver to loosen the 3 mounting screws at the bottom of the stage moving device remove the stage moving device and lay it flatly on the worktable upside down Do n...

Страница 15: ...phase contrast objective into the light path and turn the phase contrast turntable or phase contrast puller to the 10 scale Turn condenser height control knob so that the phase contrast condenser rise...

Страница 16: ...dge clear 4 In the phase contrast observation mode the center of the dark ring should coincide with that of the bright ring If their centers do not coincide as shown in Fig 20 c it is necessary to cen...

Страница 17: ...b and its surroundings have cool down This is to prevent hand getting burnt by heat 3 Unscrew the screw loosen the screw and pull out the lamp holder from the halogen lamp housing Use a cloth to remov...

Страница 18: ...nless the electric components in this unit is still operating When this unit is not to be used for a long time remove the power plug from the supply socket and keep all cables properly 2 This unit sho...

Страница 19: ...a conforming halogen lamp Optical system and imaging The nosepiece has not been turned to the fixed position Turn the nosepiece to the fixed position The filament image deviates from the center of th...

Страница 20: ...not laid flatly Lay the specimen flatly on the object stage and hold it stably The phase contrast objective does not match the ring plate of the phase contrast condenser lens Align the phase contrast...

Отзывы: