Operation and Use

 

Shipping Mode: 

ali-Q is shipped from the factory in 

Shipping

 mode for safety and to preserve battery life. When you first 

take your ali-Q out of the box, you must wake it out of shipping mode by pressing the 

aliquot button

 with 

the unit in the upright position.  

To put the ali-Q back into Shipping mode, turn the unit completely upside down and press and hold the 

aliquot button

Shipping mode should only be used when shipping or transporting ali-Q. 

Sleep Mode: 

ali-Q automatically goes into 

Sleep

 mode to preserve battery life after being motionless for 10 seconds. 

The LED will blink 

RED

 when entering sleep mode. Once the 

ali-Q

 is picked up

 or moved,

 it will 

automatically wake itself up, and LED will blink 

GREEN

Pipetting speed control:

 

In Conventional 

Pipetting 

Mode

 (see next page),

 

t

he speed that the ali-Q 

aspirates and dispenses liquid is controlled by the touch-sensitive aspirate and dispense buttons. The 
amount of force used to push on the buttons determines the speed that the liquid enters or exits the 
pipet.  

↑ BUTTON FORCE = ↑ SPE

ED 

Wall Mount:

 ali-Q may be stored in the wall mount to avoid damage from accidental drops and chemical 

spills (and save precious bench space!). ali-Q may also be placed in wall mount with 

sero 

pipet

s up to 

50mL

 installed. Wall mount will firmly adhere to any 

smooth, clean, non-painted

 surface,

 such as the 

inside of a hood

, using the 3M Command strips included

. Clean the surface with alcohol before applying 

the wall mount. Remove the stickers on the adhesive backs and immediately press onto surface and keep 
applying moderate pressure for about 30 seconds. DO NOT place ali-Q into wall mount for about 30 
minutes, to allow adhesive to set.  

<- Back to T.O.C

.

Содержание ali-Q 2000-0005

Страница 1: ...Operating Manual ali Q Aliquoting Pipet Controller Operating Manual or click here or scan QR code...

Страница 2: ...ode 7 Aliquot Mode 8 Intelligent Measuring System 9 LED Indicators 9 Technical Specifications 10 Troubleshooting 11 Maintenance 12 Cleaning Decontaminating 12 Autoclaving 12 Replacing Membrane Filter...

Страница 3: ...sories and spare parts See Ordering Information on pg 18 Do not attempt to make any technical alterations Do not dismantle the instrument any further than is described in the operating manual 9 Always...

Страница 4: ...harging port Micro USB N A 4 Aliquot button Nylon 5 Aspirate button Nylon 6 Dispense button Nylon 7 Battery Lithium ion Lithium Polymer 8 Battery compartment cover PC ABS 9 Nozzle Cover Nylon 10 Nozzl...

Страница 5: ...on verb To separate a volume of solution or suspension into aliquots Aspirate to fill with liquid Dispense to eject liquid Serological Pipet pipet disposable glass or plastic tube with volume graduati...

Страница 6: ...d control In Conventional Pipetting Mode see next page the speed that the ali Q aspirates and dispenses liquid is controlled by the touch sensitive aspirate and dispense buttons The amount of force us...

Страница 7: ...the bottom gray dispense button until bottom of meniscus is at the desired volume graduation ATTENTION Do not fill the pipet over the max volume Severe damage can occur if any liquid enters ali Q Dec...

Страница 8: ...d 4 Aliquot Carefully position the tip of the full pipet over the first receiving vessel Press and hold the purple aliquot button and the LED turns solid GREEN Hold button until green LED turns off an...

Страница 9: ...tions Temperature humidity and pressure all have significant effects on the performance of any liquid handling device EXCEPT for the ali Q ali Q can sense changes in ambient conditions and compensate...

Страница 10: ...ANY brand serological pipet Optimal performance when using VistaLab Wobble not serological pipets Not for use with Pasteur pipets Mains power supply Input Voltage 100 240 VAC Frequency 50 60 Hz Input...

Страница 11: ...Wobble not pipets will significantly decrease damage to the nozzle insert Out of Calibration Using different equipment different density liquid varying environmental conditions Membrane filter clogge...

Страница 12: ...ation lamps NOTE The effects of the UV exposure will likely cause discoloration over time Autoclaving NEVER AUTOCLAVE ALI Q Only the nozzle cover nozzle insert and membrane filter can be detached from...

Страница 13: ...cord into the charging port of the ali Q under the aliquot volume dial ali Q can s ll be used when plugged into charger Replacing the Ba ery Removing Battery Hard reset 1 Remove battery cover On the...

Страница 14: ...n at least once a year or whenever conditions and equipment change In lab Verification Adjustment Users can verify the accuracy of the dispensed aliquots in four ways 1 Gravimetric Use a digital scale...

Страница 15: ...used For warranty claims please contact Technical Support to receive Return Authorization Unit will be inspected for defects and repaired replaced as necessary at the sole discretion of VistaLab Cust...

Страница 16: ...zle Cover 1 pack 9020 9005 Replacement Power Supply w Universal outlet adapters 1 pack 9020 9008 Wall Mount 1 pack Wobble not Serological Pipets Catalog Number Description Quantity 4090 0005 Wobble no...

Страница 17: ...ervice life According to the Directive 2002 96 EC of the European parliament and of the Council on Waste Electrical Equipment WEEE of 27 January 2003 electronic equipment requires disposal according t...

Страница 18: ...General Requirements UL 61010 1 Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including...

Отзывы: