Technical information
Technical information
E
E
Error
Possible cause
Remedy
The motion control
detected movement
during the measure-
ment.
- Repeat the measurement
- Do not move your arm
- Do not speak
Measured
values/pulse
values are
extremely high
or low.
Too little rest before
taking the measure-
ment. Movement or
talking during the
measurement.
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 43.
Incorrect measure-
ment posture
Repeat the measurement and
maintain the correct posture.
(Page 47)
Measurements
taken in direct
succession
produce diffe-
rent measured
values.
Blood pressure is not
a fixed value. It can go
up or down by more
than 20 mmHg within
a few minutes.
In order to obtain compara-
ble results, always carry out
the measurements under the
same conditions.
Measurement
was interrupted,
the cuff deflated
to 0 and the
measurement
restarted.
Device detects a fault
or that the diastolic
value is too low.
Device shows no fault. Repeat
measurement
Movement during
measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes‘ rest.
No display after
device is swit-
ched on
Batteries inserted
incorrectly
Check position of batteries.
Batteries flat
Replace batteries.
Battery contacts dirty. Clean battery contacts with a
dry cloth.
Dashes in
display after
pressing Memo-
ry button
No measured values
have been stored.
Repeat measurement
Measurement is
interrupted
Batteries flat
Replace batteries.
The time is
visible in the
display, but the
measurement
is interrupted
and the symbol
appears.
Batteries flat
Replace batteries.
2. Customer service
Device repairs may only be performed by the manufacturer or an
expressly authorised body. Please contact:
Error
Possible cause
Remedy
Date and time
are not visible in
the display
Function was uninten-
tionally deactivated
Activate the date and time
function as described on
page 45.
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax.: +49 (0) 9342/924080
Email: [email protected]
Website: www.uebe.com
3. Technical data
Model:
visomat
®
comfort 20/40
Size:
L = 127.3 mm x W = 162.3 mm x H = 96 mm
Weight:
approx. 338 g without batteries and cuff
Display:
LCD display (liquid crystal display) for mea-
sured values and check displays
Memory:
2 memories each capable of storing 120
readings
Measurement method: Oscillometric determination of systole, diastole
and pulse
Clinical trial:
EN 1060-4:2004 September 2004,
Minutes of the Deutsche Hochdruckliga Ger-
man Hypertension League December 2005,
Minutes of the European Society of Hyperten-
sion July 2008,
ISO 81060-2:2009 September 2014
Reference method of
the clinical test:
Auscultatory measurement