Technical information
E
What you should know about blood pressure
D
One-off and irregular measurements therefore say little about
your actual blood pressure. A reliable assessment is only possible
when measurements are taken regularly. Discuss the measured
values with your doctor.
1. Error messages
Error
Possible cause
Remedy
Display Err - 300 Excessive cuff pres-
sure. The cuff was in-
flated to its maximum
due to the movement
of the arm or body.
The air hose may be
kinked or the air plug
blocked.
- Repeat the measurement
- Do not move your arm
- Do not speak
Inspect the hose line and air
plug for blockages
Display Err - 1
Measurement error;
measurement could
not be carried out
due to movement or
speaking.
- Repeat the measurement
- Do not move your arm
- Do not speak
The acquired values lie
outside the measuring
range of the device
(see „Technical data“
on page 59).
Repeat the measurement
Display Err - 2
Inflation error;
pressure built up too
slowly.
Check that the air connector
is fitted properly, repeat
measurement.
Fault has been iden-
tified. The cuff is not
correctly attached
- Repeat the measurement
- Do not move your arm
- Do not speak
Display 0
0
Start/Stop button was
accidentally pressed
while the batteries
were being replaced
or while the mains
adapter was being
plugged in.
Switch unit off and then on
again using the Start/Stop
button.
Arrhythmia, mo-
vement disorders,
trembling, wobbling
(objects), deep bre-
athing etc.
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device‘s
pulse beep signals. Possible
influences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed with
your doctor!