background image

DESCRIPTION

: the projection/display reads "Lo.5”.  

PROBLEM

: the room temperature is too low (below 10°C/50°F if you've pressed the 

“FACE”        button or below 5°C/41°F if you've pressed the “HOME”        button).

SOLUTION

: move to another, warmer room and, if you are taking the body 

temperature, wait for the stabilization of the device and of the subject.

DESCRIPTION

: the projection/display shows "Hi.2" and the 

value alternately. 

MEANING

: ATTENTION! The temperature is above the level 

alarm you’ve chosen (see par. 7). 

DESCRIPTION

: the projection/display reads "Hi.2".

PROBLEM

: the temperature detected exceeds the limit for operation (>42.5°C / 

>108.5ºF if you've pressed the “FACE”        button or >80°C / >176°F if you've 
pressed the “HOME”        button).

SOLUTION

: if you’ve pressed the “FACE”        button: please make certain that you’ve 

pressed the correct button, with regard to the measured object, and that the warnings 
have been met.
If you’ve pressed the “HOME”        button (DOCT mode): the temperature reading 
cannot be taken (because it is too high).

DESCRIPTION

: the projection/display reads "Lo.3".

PROBLEM

: if you’ve pressed the “FACE”        button: the forehead temperature appears 

to be too low (<34.0°C or <93.2°F).
If you’ve pressed the “HOME”        button (DOCT mode): the surface temperature read is 

below the operating limit (<1°C/33.8°F).

SOLUTION

: if you’ve pressed the “FACE”        button: make certain the waveguide  is not 

soiled or damaged and that the subject has not come from a cold room.
If you’ve pressed the “HOME”        button: the temperature reading cannot be taken.

DESCRIPTION

: the display shows "off".

MEANING

: the button that you've pressed is disabled when the device is set in NURS. 

SOLUTION

: if you want to enable it see par. 7.

12   - English -

10. TROUBLESHOOTING

The projection/display does not turn on:  the batteries are completely dead or 

incorrectly inserted; replace or reinsert them (par. #11).

The temperature is not projected between the two arches: the unit is not at the correct 

distance; move the thermometer forward or back until the projected temperature falls 

precisely between the two arches (fig. 6).

The projected temperature is not clearly visible: there is too much light in the room 

or the batteries are low; cast a shadow over the subject or replace the batteries.

The sensor (fig. 1) is damaged or water has seeped into the thermometer: contact 

your Dealer right away for technical service.

The thermometer temperature reading is too low:

- make certain that the conditions outlined in the warnings (par. #2) have been met;

- check that the waveguide (fig. 1) is not soiled or damaged; if it is, clean it as 

indicated in par. #12 or contact your Dealer for technical service;

- check that the thermometer is perpendicular to the forehead as indicated in fig. 3.

The temperature reading is too high: make certain that the conditions outlined

1.

2.

3.

4.

5.

6.

Содержание 06480

Страница 1: ...mometer Mode d Emploi Lire attentivement les instructions avant d utiliser le thermom tre Seguire le istruzioni per l uso Follow instructions for use Suivre les instructions d utilisation Riferirsi al...

Страница 2: ...ttery door 4 x AAA with brief instructions Slide to open Special lanyard with additional brief instructions to be worn around the neck Faisceau lumineux de positionnement Orifice avant Capteur au fond...

Страница 3: ...ry to wait the end of the countdown without touching the device Alternatively at any time it is possible to do the manual MQCS suggested procedure In DOCT mode the HOME and the Mem buttons are enabled...

Страница 4: ...e projection on the same spot precisely at the center of the forehead midway between the top of the nose and the hairline and keep the thermometer perpendicular to the forehead Do not take measurement...

Страница 5: ...e mentioned precautions par 2 1 may lead to very low or very high temperature readings which cannot be attributed to product s malfunctions The aiming lights meet the photo biological safety requireme...

Страница 6: ...l temperature varies according to the measurement s site and throughout the day also in response to physical or mental effort for example a baby s crying Moreover the body temperature can be affected...

Страница 7: ...the HOME button release the button lights will flash twice s l o w l y and then the display will show the acquired temperature VisioFocus PRO is now ready to take a measurement For more details about...

Страница 8: ...econds it reverts to stand by mode and will display the room temperature for 45 minutes DOCT mode or 30 minutes NURS mode before shutting off 4 2 In case of sweating oxygen mask elderly patients take...

Страница 9: ...any other problem which may cause an alteration of the surface skin temperature Of course VisioFocus PRO can be used on open wounds or organs during surgical interventions ensuring maximum hygiene th...

Страница 10: ...your thermometer leaves the factory Celsius or Fahrenheit degrees referred to oral rectal or axillary or internal temperature readings or only internal temperature readings no symbol will be shown NUR...

Страница 11: ...h air conditioning in the room when appears wait for and release the button the display will show alternately the word and the temperature value otherwise choose 8 ROOM TEMPERATURE CALIBRATION 8 1 MQC...

Страница 12: ...e MQCS can also be performed in case of countdown see par 8 2 In NURS setting the MQCS is mandatory every 30 minutes In DOCT setting see the par 8 2 8 2 AQCS countdown only DOCT MODE In DOCT setting i...

Страница 13: ...he thermometer has not stabilized SOLUTION wait until the AQCS countdown has run its course without touching the thermometer or perform an MQCS par 8 DESCRIPTION while in stand by mode the room temper...

Страница 14: ...ressed the FACE button make certain the waveguide is not soiled or damaged and that the subject has not come from a cold room If you ve pressed the HOME button the temperature reading cannot be taken...

Страница 15: ...king a temperature reading or run a manual quick calibration MQCS par 8 Remove the batteries if you do not expect to use the thermometer for a long time 12 CLEANING CLEANING THE WAVEGUIDE the thermome...

Страница 16: ...the operating range are not guaranteed and the message Hi 4 and the temperature value are displayed alternately Power supply 4 AAA LR03 alkaline batteries 1 5 V included Life of high quality batteries...

Страница 17: ...metro perpendicolare al centro della fronte Correct distance and thermometer perpendicular to the forehead Bonne distance et thermom tre perpendiculaire au milieu du front Fig 3 Fig 6 Distanza corrett...

Отзывы: