Operating instructions, Quick installation
12
7. Edit the analogue output type.
The signal type for the analogue output is defined by parameter ‘Ana. output type’
that has a number of alternatives.
Press
+
to get parameter ‘Ana. output type’ displayed.
Ana. output type
4-20mA
Press
if the parameter should be edited, and make the editing
like in point
3.
on page 10.
8. Chose a calibration type.
Press
+
to get the parameter ‘Calibration type’ displayed.
Calibration type
Data sheet
The value points out which type of calibration that was performed last time.
Two calibration types can be chosen in ‘Quick set-up’: Data sheet and Deadweight.
As a new calibration shall be performed, WST 3 must first be set for editing, indicated
by a cursor at the parameter value. Then one of the alternatives must be chosen and
accepted.
If a new calibration shall be performed, press ENTER to switch the transmitter
into set-up mode.
Calibration type
Data sheet
A cursor starts flashing to the left on the parameter value line.
Chose an alternative for the parameter like in point
3.
on page 10.
When
is pressed for 2 seconds, the chosen calibration type gets active and
the calibration can start.
The choice of calibration type determines which parameters will follow:
For Data sheet, se point
9.
on page 13.
For Deadweight, se point
9.
on page 16.