Operator’s Manual
Operating Procedure
29/50
On the Running Screen a schematic representation of the instrument’s layout is displayed.
On the left the Samples Rotor
is shown,
on the right at the bottom the Strip Loading Rotor is shown, on the top, the
Reaction Rotor. At the right side of the Strip Loading Rotor the Reagent’s
Rack is present.
On the Left side of the Screen three information Tabs are displayed:
•
Strip rotor Clears in: time needed to have the Strip rotor available to load other strips, at start up its
status is Free.
•
Strip Positions available: number of available positions on the Strip
Rotor (39 in the empty position is in
the strip rotor/40 if the empty position is in the reaction rotor)
•
Next Loading Phase in: time needed to be waited before to be able to load other Samples and/or Strips,
at start-up it is now.
In the Bottom Bar of the program Window there are the buttons that allow to:
•
Refill the Wash Buffer Tank: Refill Buffer
•
Empty the Waste Tank: Empty Waste
•
Perform the Daily Maintenance: Maintenance.
To start the Session Programming press Add Samples.
3.3.4.
L
OAD
S
AMPLES
S
CREEN
The Load Samples screen appears, showing the free positions on the sample rotor where the sample’s tubes can be
placed.
Insert the sample tubes in the available positions with the barcode label facing toward the outside of the rotor (see
picture below).
For each tube the dimensions in the corresponding row of the table must be set, the eligible dimensions could be
selected using the drop-down menu of each row.