Operator’s Manual
Operating Procedure
23/50
•
Before any connection, ensure there is a proper grounding terminal.
•
Avoid the use of extension or movable cord.
WARNING: All the repairs must be carried out by the VirCell S.L. trained staff. Do not open the electronics board, on
the front left part of the instrument, without having first disconnected the power cords.
WARNING: The Manufacturer doesn’t guarantee the safety and performance requirements in case the VirClia Lotus is
powered with a cord that is compatible with the voltage of the country where the instrument is installed but other
than the one supplied by the manufacturer.
•
In case of malfunction or evident damages, disconnect the power cord and contact the Technical service of local
distributor.
•
Use the instrument only as described in this Manual of VirClia Lotus. Any use other than the one for what the
instrument is intended, is to be considered improper and can reduce the protection offered by the instrument.
•
The Manufacturer cannot be held liable for damages deriving from an improper, incorrect and / or unreasona-
ble use of the instrument or if the instrument is used in electric systems that do not comply with the current
safety regulations of the country where it is installed.
Environmental Requirements
Install VirClia Lotus on a level surface in a place where a constant temperature and a proper level of humidity are
guaranteed. Avoid direct exposure to sunlight. Connect the VirClia Lotus plugs to the power outlets that are grounded.
Supply power VirClia Lotus only with voltages included in those set by the technical characteristics of the instrument.
All these requirements are stated in the technical specification section.
Waste Disposal
Disposable accessories and consumables used and that may be contaminated must be treated as biological waste in
accordance with the procedures established by the Laboratory and collected by a specialized waste manager. All ma-
terials used should be handled and disposed of as potentially infectious. Observe local regulations for the disposal of
clinical waste
Preparation of Sample
Verify that there are no clots, foam or bubbles in the samples prior to runs despite VirClia Lotus can detect their
presence.
NOTE: The information about the preparation of samples, provided in this Manual, is of general nature. The operator
must carefully read the instructions for use contained inside each kit.
WARNING: An improper preparation of samples can lead to false negative results. Samples potentially containing
particles must be centrifuged prior being used.