ED12DSS User Manual
9
Getting Started
Features
Open-frame printed circuit card suitable for OEM applications.
RoHS Compliant
Operates DSS or NS series bi-stable shutter devices.
2-Pin JST shutter interface connector
203D
Shutter interconnect cable included – JST 2-pin to JST 2-pin, 1 meter in length
Exposure determined by external pulse (BNC,TTL active-high) or switch contact closure when
switching in5 VDC into trigger input
In-line PULSE INPUT BNC active-high +5.0 V TTL
Selectable pulse high-current duration for specific DSS/NS series shutters via 4-position piano
switch
Internal fuse protection for SHUTTER OUTPUT
Operates on +12 to +24 VDC at 1.5 A
Power input to controller via 2-wire input harness
ED-IOP
Input cable harness included for power and trigger signal input.
+5.0 V TTL signal input (BNC)
+5.0 VDC output to enable driver from external switch contact or transistor switch
closure.
Input/output interface cable
18-inch red/black bare power leads
6-inch BNC trigger input with reference ground.
Both terminate to Tyco/Amp 5-pin polarized connector for connection to
ED12DSS.
Re12 to +24 VDC External power supply (user supplied) for operation
Size: 0.50 x 2.25 x 2.25 inches (12.7 x 57.2 x 57.2 mm)
Weight: 0.730 ounces (0.021 Kg)