ED12DSS User Manual
11
Start Up
After unpacking your unit inspect for any defects. Upon inspection if a visible defect is found, or a part
(or parts) is missing, notify Vincent Associates immediately.
Prior to the connection of input/output signals to the
ED12DSS
be sure that:
1.
The
5-pin connector of the
ED-IOP
(power/trigger input connector) is disconnected from the 5-
pin post connector P1 of the
ED12DSS
.
2.
The power supply is in the off position.
3.
Set S1 to the pulse duration required for the shutter being used. See
SPECIFICATIONS
and
Figure 1 elsewhere in this manual for required settings.
4.
Connect the Red wire of the
ED-IOP
harness to the (+) side of a +12 VDC to +24 VDC 1.5A
power supply.
5.
Connect the Black wire of the harness to the (-) return side of the power supply.
6.
Connect the 5-pin female connector of the
ED-IOP
onto P1 of the
ED12DSS
. (The connector is
polarized and can only be properly connected in one orientation)
7.
Connect the
203D
shutter interconnect cable between the 2-pin P2 connector on the
ED12DSS
to
the 2-pin connector on the shutter’s flex interconnect.
8.
Turn on the power supply.
9.
Connect an input signal, active-high TTL to the input BNC.
Be sure to observe Electro-Static Discharge (ESD) anti-static
unpacking procedures when removing the ED12DSS from the
static shielding bag. Improper handling can result in
destruction of the CMOS integrated circuits on the board.
CAUTION