background image

TS21I_IR_CO2 User Manual 

Vighnaharta Technologies Pvt Ltd                                      Page 

9

 of 

13

 

 

 

 

1008 

%v/v Higher 16-bit 

 

 

1009 

%v/v Lower  16-bit 

1002/2 

  Read  Threshold Low 

  

1003/3 

  Read Threshold High 

1002/2 

  Read  Threshold Low  

  

1003/3 

  Read Threshold High  

101 

Write D02 High Alarm 

  

102 

write D01 Low Alarm 

1002 

Write Threshold Low  

  

1003 

Write Threshold High  

 

Understanding controls on front panel 

 

LED Indicators 

1. 

ON 

Power 

ON

 LED. Green LED turns on when power is ON. It 

indicates Power On status of the unit. 

2. 

L – Alarm 

When gas concentration exceeds low threshold level, 
TS21I_IR_CO2 generates low alarm condition. For low alarm 
condition, yellow LED turns ON till unit is reset.If relay1 output 
is pulse type, relay remains on for 5sec. If relay1 output is 
latch type, relay remains on till the unit is reset.  

3.  

H- Alarm 

When gas concentration exceeds high threshold level, 
TS21I_IR_CO2 generates high alarm condition. For high alarm 
condition, red LED turns ON till unit is reset. If relay2 output is 
pulse type, relay remains on for 5sec. If relay2 output is latch 
type, relay remains on till the unit is reset.   

3.  

F- Alarm 

When supplay goes below 19V or above 29V and also sensor 
status open or short. Orange LED turns ON till unit is reset. If 
relay3 output is pulse type, relay remains on for 5sec. If 
relay3 output is latch type, relay remains on till the unit is 
reset.    

4. 

 LED Display  
 
 

Four  digits  -  Seven  Segment  LED  display  shows  the  current 
gas concentration in %v/v. Also it shows configuration settings 
like current / voltage range.  
Example, the seven segments LED display shows the 0.046 
%v/vof gas concentration. 

Magnetic Keypad - 

Four magnetic keys operate using magnetic stylus. 

1. 
 

 ENTER       

Move stylus to 

ENTER

 key to enter into system to access 

menus. This key has dual function.  It is used to Enter or Save 
the data. User can select some menu or can save settings 
made by the user. If user presses the key during Idle mode, it 
displays the menu.   

2. 

 ESCAPE 

Move stylus to 

ESCAPE

 key to exit from the current menu. If 

user exits without saving, current settings will not be saved.   

3. 

 UP 

Move stylus to 

UP

 key to move menu up side or increment the 

menu. 

4. 

 DOWN 

Move  stylus  to 

DOWN

  key  to  move  menu  down  side  or 

decrement the menu. 

 

 

0 .0  4  6 

Содержание TS21 I IR CO2

Страница 1: ...n Website http www vighnaharta in Prepared By MDC Approved By JG Doc No TS21I_IR_CO2_UM_2_0 Rev No 2_0 Eff Date 23 05 2020 TS21I_IR_CO2 USER MANUAL Mfg By Vighnaharta Technologies Pvt Ltd 2nd Floor Karanjkar Estate S N 11 11 Nanded Gaon Pune 411041 Contact no 919011033259 Landline Sales 919011033260 Email sales vighnaharta in Website http www vighnaharta in ...

Страница 2: ...21I_IR_CO2 User Manual Vighnaharta Technologies Pvt Ltd Page 2 of 13 Table of Contents Overview III Technical Specifications IV Installation Procedure IV Operation VI Menu Details X Calibration process XI ...

Страница 3: ...s CO2 sensors can be used to monitor the quality of air and the tailored need for fresh air respectively Measuring CO2 levels indirectly determines how many people are in a room and ventilation can be adjusted accordingly Salient Features Detection Gas CO2 carban dioxide Gas Detection Range v v 0 to 5 v v with user defined high and low alarm level Four digit seven segment LED display for v v level...

Страница 4: ...w driver Allen Key set To install the unit at the user premises 1 Run all desired power siren and hardware contact wires to the panel location 2 Level the cabinet on sturdy wall and mark the mounting holes with a pencil Installation precautions This detector head in the flame proof enclosure must be installed near the CO2 gas cylinders or at the joints of gas carrying pipes The detector must face ...

Страница 5: ...NO Connector CON2 Pin No Description 1 D RS 485 data 2 D RS485 data 3 C ve current Output 4 C ve current Output 5 A O P ve analog output voltage 6 A O P ve analog output voltage 7 DI Digital Input 8 VCC 9 GROUND Connecting wires Take out the wires from supply relay output and RS 485 port Make the connections for relay and RS 485 as mentioned in above table Take out wires from output and connect it...

Страница 6: ...es four magnetic keys which operate using single magnetic stylus to do settings like low high alarm levels relay output type and so on Front panel of typical TS21I_IR_CO2 is shown as below Relay Settings TS21I_IR_CO2 has threerelay outputs Relay1 for low alarm condition Relay2 for high alarm condition and Relay3 for Fault operation User can use it as a NO or NC type output using the respective jum...

Страница 7: ...n set NO type contact or NC type of contact For Relay1 JP1 Jumper between pin 1 and 2 Relay1 will be NC type Jumper between pin 2 and 3 Relay1 will be NO type JP1 For Relay2 JP2 Jumper between pin 1 and 2 Relay2 output will be NC type Jumper between pin 2 and 3 Relay2 output will be NO type JP2 For Relay3 JP3 Jumper between pin 1 and 2 Relay3 will be NC type Jumper between pin 2 and 3 Relay3 will ...

Страница 8: ...slave address using Interaction Maximum of 254 slave IDs can be set using Interaction Refer Interaction details Note If user set wrong slave address then unit may not be able to communicate with gas detection panel Modbus Communication Default Settings for TS21I_IR_CO2 Baud Rate 9600 bps Parity Even Relay 1 NO O P PULSE type Relay 2 NO O P PULSE type Relay 3 NO O P PULSE type Output Voltage 0 10Vo...

Страница 9: ...y remains on for 5sec If relay2 output is latch type relay remains on till the unit is reset 3 F Alarm When supplay goes below 19V or above 29V and also sensor status open or short Orange LED turns ON till unit is reset If relay3 output is pulse type relay remains on for 5sec If relay3 output is latch type relay remains on till the unit is reset 4 LED Display Four digits Seven Segment LED display ...

Страница 10: ...H type 1 Press ENTER to go in main menu 2 Use Down Key to select oP L function 3 To go into oP L function again press ENTER key 4 Use Up and Down keys to select PULSE or LATCH type for Relay1 output 5 Press Enter key to save the settings or press Escapekey to return to previous menu without saving the settings 4 oP h Output High User can set Relay2 output as either PULSE or LATCH type 1 Press ENTE...

Страница 11: ...ress ENTER key to save the count 9 Press ESCAPE key to exit from Configuration mode and remain in Idle or Scan mode 2 o2ZC Output 2 Zero calibration This is a calibration function used to calibrate voltage output for 0V o2SC Output 2 Span calibration This is a calibration function used to calibrate voltage output for 5 10V depending on voltage range set in menu A oP 1 Go to menu A oP 2 Press Up or...

Страница 12: ...thout saving the settings Do s and Don ts when gas leak occurs Do s If the gas leakage is severe it is very dangerous for anybody to enter the premises as it causes shortness of breath headache nausea dizziness etc Call Gas dealer turn ON exhaust If there is a minor moderate gas leakage Turn off the knob regulators etc to stop gas supply Open all doors windows Don ts Do not panic Do not OFF ON any...

Страница 13: ...nnection therewith No statement or recommendation contained herein shall have force or effect unless in an agreement signed by officers of manufacturer or seller Seller s and manufacturer s only obligation shall be limited to replace the product which is proved to be having manufacturing defect and returned with in the warranty period Terms Of Warranty 1 All models of TrueSafe detectors are warran...

Отзывы: