TS21I_IR_CO2 User Manual
Vighnaharta Technologies Pvt Ltd Page
12
of
13
Function names in the menu
Process Steps
from
following
option
2.4(2400),
9.6(9600),
19.2(19200).
5.
Press
Enter
key to save the settings or press
Escape
key to return to previous menu without
saving the settings.
4.
PArt
User can set Parity either None, Even, Odd.
1.
Press
ENTER
to go in main menu.
2.
Use
Down
key to select ‘
PArt
’ function.
3.
To go into ‘
PArt
’ function, again press
ENTER
key.
4.
Use
Up
or
Down
keys to select any parity from
following option ‘
nonE
’ (None), ‘
EuEn
’ (Even),
‘_
odd
’ (Odd).
5.
Press
Enter
key to save the settings or press
Escape
key to return to previous menu without
saving the settings.
5.
Addr
User can set device address from 1 to 254.
1.
Press
ENTER
to go in main menu.
2.
Use
Down
key to select ‘
Addr
’ function.
3.
To go into ‘
Addr
’ function, again press
ENTER
key.
4.
Use
Up
or
Down
keys to select address from 1 to
254.
5.
Press
Enter
key to save the settings or press
Escape
key to return to previous menu without
saving the settings.
Do’s and Don’ts when gas leak occurs
Do’s:
•
If the gas leakage is severe, it is very dangerous for anybody to enter the premises, as
it causes shortness of breath, headache, nausea, dizziness etc. Call Gas dealer/ turn ON
exhaust.
•
If there is a minor/moderate gas leakage- Turn off the knob regulators etc, to stop gas
supply.
•
Open all doors & windows.
Don’ts:
•
Do not panic.
•
Do not OFF/ON any electrical switch as CO2 is readily flammable gas.
Routine Check for TrueSafe TS21I_IR_CO2
•
To ensure that TrueSafe TS21I_IR_CO2 is in working condition, it is necessary to check
working of Gas LED by manual testing.
•
For checking unit’s sensing capability, it is necessary to check the unit once in a month.
Maintenance
Maintenance Guidelines
:
• The sensor is shipped with a factory calibration. Sensor should be calibrated 6 months from
purchase date.
• Calibrate the detector at least once every 1year.
• Calibration should be performed with certified calibration gas. Calibration kits and replacement
cylinders are available from Calibration Technologies.
• All tests and calibrations must be logged.
• Always disconnect power at the controller before performing any wiring at the sensor.