CAUTION FOR USE
Cautions are intended to alert you to importance of following the correct operating
procedures where risk of injury to the patient or system.
Do not handle the plug with wet fingers ( To avoid electric shock)
The operators are here by instructed not to touch the equipment and the patient
simultaneously.
GENERAL SPECIFICATION
Equipment classification
Mode of operation
: 5 minutes (Max)
Degree of mobility
: Portable
Type of protection against
Electrical shocks - Class 1
Degree of protection against
Electrical shocks - Type ‘B’
Power Requirements
AC Input I/P: 100-230V AC, 50-60Hz,
O/P: 5V DC, 1A
Fuses
1A Slow Blow (110V AC)
500mA Slow Blow (230V AC)
3A Slow Blow
Safety Rules:
The series family ultrasound machines are designed to comply with the following
safety aspects.
x
Leakage current to ground less than 0.5mA in normal condition.
x
Leakage current to ground less than 1.0mA in single fault condition.
x
Isolation voltage between ground and mains wires greater than 1500V.
x
Emission and susceptibility requirements for class A. equipment as per IEC 60601-1-
2 standards.