19440-001 Rev. E
VIDAR NDTPRO Industrial Film Digitizer Installation Guide
30
Guidance And Manufacturer’s Declaration – Electromagnetic Immunity
Enclosure Port
The VIDAR NDT
PRO
Industrial Film Digitizer is intended for use in the electromagnetic environment specified below. The
customer or end user of the Digitizer should assure that it is used in such an environment.
Immunity test
Test method
Immunity Test Levels
Professional healthcare facility environment
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ±4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
See Table A
Rated power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 or 60 Hz
Guidance And Manufacturer’s Declaration – Electromagnetic Immunity - Input a.c. power
The VIDAR NDT
PRO
Industrial Film Digitizer is intended for use in the electromagnetic environment specified below. The
customer or end user of the Digitizer should assure that it is used in such an environment.
Immunity test
Test method
Immunity Test Levels
Professional healthcare facility environment
Electrical fast transient /
bursts
a) l) o)
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surges
a) b) j) o)
Line-to-line
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Surges
a) b) j) k) o)
Line-to-ground
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ±2 kV
Conducted disturbances
induced by RF fields
c) d) o)
IEC 61000-4-6
3 V
m)
0,15 MHz – 80 MHz
6 V
m)
in ISM bands between 0,15 MHz and 80 MHz
n)
80 % AM at 1 kHz
e)
Voltage dips
f) p) r)
IEC 61000-4-11
0 %
U
T
; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
q)
0 %
U
T
; 1 cycle
and
70 %
U
T
; 25/30 cycles
h)
Single phase: at 0°
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
0 %
U
T
; 250/300 cycle
h)
Note:
U
T
is
the A.C. mains voltage prior to application of the test level.