12
MIN/MAX/AVG
Relative Base value
0
0
Frequency of Power
50Hz
Changeable
Back Light Time
0000 S
Changeable
Continuity beep
when the input drops below 150
Ω
Changeable
Compare beep
if the reading lower or higher set value
Changeable
Changing the Power-Up Configuration
You can change the power-up configuration to one that more closely meet your needs and preferences.
To change the Power-Up Configuration, power-up the meter and press SHIFT simultaneity
1. Change Back Light Time: Press AUTO, when “bLOFF” is annunciator showed on the primary
display. The value of setting is showed on the secondary display, then press
or to step up or down the setting value with 30S. The limit of the setting value is
3600S(1 hour), when the setting value is “0000” indicates that back light only be turned off
manually. Press the HOLD save the set value, the “memory” annunciator is lit on.
2. Change Frequency of Power: Press AUTO, when “FrSET” annunciator is showed on the primary
display, the Frequency is showed on the secondary display, then press or
to select the frequency at 50Hz or 60Hz. (Once change the Frequency of Power, the meter
should be calibrated newly). Press the HOLD save the set value, the “memory” annunciator is lit
on.
3. Change the Reference Impedance: Press AUTO, when “CONS” annunciator is showed on the
primary display, the Reference Impedance is showed on the secondary display. Then press
or to select the needed value (supplied value are 50, 75, 93, 100, 150, 300, 500, 600, 800,
1200, 2400 (unit:
Ω
)). Press the HOLD save the set value, the “memory” annunciator is lit on.
4. Change the continuity beep mode: Press AUTO, when “CH-BB” annunciator is showed on the
primary display, the “YES”(connect beep) or ”NO”(disconnect beep) is showed on the secondary
display. Press the HOLD save the set value, the “memory” annunciator is lit on.
5. Change the compare beep mode: Press AUTO, when “CO-BB” annunciator is showed on the
primary display, the “YES”(inside the limiter) or ”NO”(exceed the limiter) is showed on the
secondary display. Press the HOLD save the set value, the “memory” annunciator is lit on.
6.
Factory set: Press AUTO, when “FACTY” annunciator is showed on the primary display, select
the default factory set. To resume the factory set, Press the HOLD save the default value. To exit
factory set mode, power off the instrument.
SECTION 4
Maintenance
NOTE
When servicing the meter, use only the replacement parts specified
INTRODUCTION
Section 4 provides the information necessary for the user to perform basic maintenance. Users should
not attempt to perform maintenance not described in this section. Qualified service personnel should
refer to this Manual for complete maintenance, service.
CAUTION
TO AVOID ELECTRICAL SHOCK OR DAMAGING THE METER, NEVER GET WATER INSIDE