background image

 

18 

Changing the sampling number, if necessary. 









数値の変わり方

            

   Pcs

M

    





How to change the value

 

If samples to be counted widely deviate 
in weight, or a higher measure of 
accuracy is desired, it is recommended 
that users change the sampling number 
to a larger value. 

Press the  Zero/Tare  key. 

Each press of the key changes the value 
on the right end.    Select the desired value. 
If the sampling number need not be 
changed, go on to the next step. 

5 Loading 

samples 

            

   Pcs

M

    





 

Load the number of samples displayed.   
Count the samples precisely and load 
them in the center of the measurement 
pan. 

Saving the unit weight of samples 

            

   Pcs

       



 

Press the  Memory  key. 

The balance saves the unit weight and 
reverts to measurement mode. 

  Key Points of the Procedure 

 

1. 

While the samples are being saved, the value indication disappears and only [M] flashes to 
indicate that memory saving is underway.    If the balance is affected by wind or other vibrations 
during this process, the saving time may be prolonged. 

2. If 

[



] appears, it indicates one of the following states: 

(1)  The weight of one sample (measurable unit weight) is insufficient. 

For the range of unit weights that can be measured and saved, please refer to “11.   
Specifications,” on page 30. 

(2)  In the sampling of Operation Step 3, press the    Set    key with the samples loaded on the 

balance.  

 If [



] appears, the sampling is interrupted and the data in progress is not saved.   

3. 

The operation for increasing counting accuracy is referred to as the Memory Update Method.   
This procedure updates the memory with a unit weight that represents a more precise average 
by gradually increasing the sampling number.   
This operation improves counting accuracy and is recommended for the following cases; 

(1)  When the samples to be counted deviate widely in weight or the number of samples 

displayed deviates. 

(2)  When greater accuracy is desired. 

4. If 

[



] appears in Memory Update Method, it indicates that a counting error is likely due to 

the small number of the samples loaded on the balance.    [

] will light up at the judgment 

indication “LO.” As the memory update continues, counting accuracy improves and the above 
indication disappears. 

5. 

If you change the sampling number, subsequent sampling will start from the new sampling number.   

Содержание SJ-1200CE

Страница 1: ...Balance SJ CE Series Operation Manual To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe place near the balance so you can rev...

Страница 2: ......

Страница 3: ...keys Furthermore the large liquid crystal display provides excellent visibility and the instrument s high speed and stability intrinsic to a tuning fork design help boost operational efficiency Before...

Страница 4: ...surement 15 6 Counting Pieces 17 6 1 Sampling 17 6 2 Increasing the Counting Accuracy Memory Update Method 19 7 Measuring Percentage 20 8 Limit Function 22 8 1 Limit Function Setup 22 8 2 Setup of Lim...

Страница 5: ...ENDED This term indicates steps that the user should take to ensure the quality and reliability of the balance Meanings of Symbols Each symbol is accompanied by an instruction Mandatory Symbol Indicat...

Страница 6: ...ric shock Keep Dry Do not use the balance in a location were it may be subjected to excess moisture Electric shock or short circuiting could occur The balance may be corroded with resultant malfunctio...

Страница 7: ...increase in the balance may lead to inaccurate measurement Unplug Adapter If the balance is to be unused for an extended period of time unplug the adapter This conserves power and prevents deteriorati...

Страница 8: ...g the level of the balance Turn the adjusters until the bubble rests in the center of the red circle Adjusters Round pan type Total of two one on either side of the unit s front Square pan type Total...

Страница 9: ...hile span adjustment is in progress Auxiliary scale interval Lights up only when the auxiliary scale interval is displayed 1 Bar graph ct carat 3 oz ounce lb pound ozt troy ounce dwt penny weight tl t...

Страница 10: ...mit function Function key Brief press toggle switches the units to be displayed in succession g Pcs etc Brief press moves the flashing digit in the setup of a limit value when using the value input me...

Страница 11: ...in the level Turn the adjusters until the bubble rests in the center of the red circle on the level Caution Use caution when operating the adjusters on the square pan type to prevent them from liftin...

Страница 12: ...mething is loaded the display does not change after In this case remove the load 3 Switching the measurement mode Press the Function key Each time the key is pressed the unit of measurement changes as...

Страница 13: ...he weight of the sample loaded in the tare Weighing only the weight of an added sample 4 Resetting the indication to zero 0 0 F NET g Press the Zero Tare key The total weight of the sample loaded on t...

Страница 14: ...er vibrations Use the windshield or vibration dampers to protect against such adverse effects Unstable Stable 0 F g 0 F g 4 When the read out indicator is reset to zero or the tare is subtracted the b...

Страница 15: ...2 Selecting the next item Press the Function key The indication changes to the next item Limit function 3 Selecting an item Pressing the Function key advances the function items at the rate of one ite...

Страница 16: ...imit values are set up judges among HI OK and LO Disable Auto zero zero tracking Enable This function automatically sets the zero point exactly to zero to prevent slight deviations Disable balance ope...

Страница 17: ...200CE when unsealed 4 3 Interface Section Displayed when is set to or Item Set Value Description Stop output Output continuous at all times Output continuous if stable stop output if unstable Outputs...

Страница 18: ...available in SJ 6200CE SJ 12KCE 5 2 Setup of Units of Measurement Only when the balance is not sealed When values to are entered prior to use the desired unit of measurement to be displayed can be cho...

Страница 19: ...ed simply pressing the Function key For more information on the units of measurement that can be set please refer to 4 2 Description of Functions on page 13 2 The units are displayed in the same seque...

Страница 20: ...Sample unit weight weight unit Sample weight Total g g 1 500 500 pcs Measured number of samples If samples to be counted deviate widely in weight or a higher measure of accuracy is desired it is reco...

Страница 21: ...ollowing states 1 The weight of one sample measurable unit weight is insufficient For the range of unit weights that can be measured and saved please refer to 11 Specifications on page 30 2 In the sam...

Страница 22: ...ples Pcs M Add approximately an equal number of samples as the number currently displayed Add them gradually in groups of approximately the same number until the number of samples is approximately dou...

Страница 23: ...ce weight 1 Activating the percentage mode Press the Function key to disaplay 2 Setting the reference value Press the Set key The display begins flashing The balance is now ready for reference value s...

Страница 24: ...The weight of the reference sample is insufficient For the limit weight that can be saved limit weight please refer to 11 Specifications on page 30 2 While setting up the reference value in Step 2 th...

Страница 25: ...s the limit value 2 Numeric value setup method The limit value is entered with a key stroke The limit values entered are held in memory even when the balance is powered down The respective limit value...

Страница 26: ...ion Press the Zero Tare key to select the desired condition 5 Setting the judgement range Judge entire range Judge beyond 5 Press the Function key The display changes to Judgement Range Press the Zero...

Страница 27: ...emory key After the lower limit value has been saved the balance displays it briefly and proceeds to the following setup If One point setup was chosen the setup is complete 4 The upper limit value set...

Страница 28: ...ess the Function key The flashing moves to the digit on the immediate left Each time the key is pressed the flashing digit moves one position left When the leftmost digit is selected the flashing adva...

Страница 29: ...imit value displayed Measurement mode restored Continuous pressing g M g M g 2 If you make a mistake press the Function key during the setup of actual quantities or the Set key during the setup of val...

Страница 30: ...ent should be performed with the balance installed perfectly level and without any load on the measurement pan 1 Activating the span adjustment function Pressed continuously Key released Press and hol...

Страница 31: ...ity of the balance To implement a calibration as precisely as possible use a weight close to the capacity of the balance Calibration weights can be ordered from Shinko For ordering information please...

Страница 32: ...rent when the power is switched on Overcharging will also decrease battery life 2 When the balance is used for the first time after purchase the operating time may be shorter than when using a fully c...

Страница 33: ...insufficient Execute the Memory Update Method 19 appears before the capacity is reached Gross weight exceeded the capacity of the balance weight range container weight of sample A section of the mech...

Страница 34: ...40 180 160 12 2 Common Specifications 1 Weight measuring method Tuning fork vibration method 2 Tare subtraction range Total capacity 3 Liquid crystal display LCD Seven segments two segments in leading...

Страница 35: ...01 7 13 19 38 70 130 190 380 troy ounce ozt 0 001 0 001 0 001 0 01 0 01 0 01 0 01 0 01 140 270 390 770 1400 2700 3900 7700 penny weight dwt 0 01 0 01 0 01 0 1 0 1 0 1 0 1 0 1 5 8 11 16 32 58 110 160 3...

Страница 36: ...aiwan 37 5 187 5 1 32277 0 08267 1 20565 24 11306 1mom 3 75 18 75 0 13228 0 00827 0 12057 2 41131 1to 11 66380 58 31902 0 41143 0 02571 0 37500 7 5 unit tael Hong Kong tael Singapore Malaysia tael Tai...

Отзывы: