12.4.5
Judgment result when the limit function is enabled (S1: 1 character)
S1
Code
Description
L
4CH
Too little (LO)
G
47H
Proper (OK)
H 48H
Too much (HI)
1- or 2-point scale
1 31H
Rank 1
2 32H
Rank 2
3 33H
Rank 3
4 34H
Rank 4
5 35H
Rank 5
3- or 4-point scale
T
54H
Cumulative value
U
55H
Unit weight
(SP)
20H
No judgment result or no data type
specified
d
64H
Gross
Data type
12.4.6 Status
(S2:1
character)
S2
Code
Description
S
53H
Data stable *
U
55H
Data unstable *
E
45H
Data error (Indicates that data other than S2 is invalid.)
(
[
], [
]
)
(SP)
20H
No status specified
* Data that is independent of whether the weighing condition is stable or not, such as cumulative
values and unit weights, is independent of whether S2 is S or U when it is output.
12.4.7
Output data other than measurement data
(1) Interval
output
When interval output is started or stopped, a header and a footer are output.
Header
- is output for 15 characters.
1 2 3 4 5 6 7 8 9 10
11
12
13 14 15
- - - - - - - - - - - - - - -
Footer
Two linefeeds are inserted.
(2) Time
output
When the time stamp function is enabled, the time is output one line above output data.
1 2 3 4 5 6 7 8
h
h
: m
m
: s
s
* hh: hour (00-23), mm: minute (00-59), ss: second (00-59)
48