4-1 SPAN CALIBRATION (CAL)
SECTION 4
ADJUSTMENTS AND SETTINGS
1. AJ Type
1. It is recommended to use a calibration weight of better
accuracy than the scale.
2. The calibration is available with 1/2 of F.S. Nevertheless, it is
recommend to us weight closer to F.S. for accurate calibration.
3. Error messages:
n
: The calibration weight is over the full capacity
b
n
: The calibration weight is less than 1/2 of the capacity.
c
n
: The data error exceeds 1%. Or perhaps the scale
may be defective.
8