37
9.5.2 Command format
(1) Command
format
Composed of four characters (ASCII), including the terminators (CR=0DH, LF = 0AH)
1 2 3 4
C1 C2 CR LF
(2)
Instruction for zero adjustment
C1 C2
Code
Description Value
Response
T (SP)
54H 20H
Instruction for
zero adjustment
None
A00: Normal end
E01: Zero adjustment cannot be
executed due to an error in
the weight value.
(3) Setup of output control
C1 C2
Code
Description
O
0 4FH
30H
Stop
output
O
2
4FH
32H
Output continuous if stable (stop output if unstable)
O
4
4FH
34H
Outputs once if stable. Outputs if the balance is stable when
a sample is loaded after the preceding sample has been
removed and the balance indicated zero, or less.
O
5
4FH
35H
Outputs once if stable, and stops output when unstable. Even
if the sample is not replaced, the balance is output once when
it stabilizes next time (including the zero indication).
O
7 4FH
37H
Pressing Memory key causes the balance to output once
when stable.
O
9
4FH
39H
Output once after stabilization.
※
The output controls executed with commands [O0] - [O7] work the same as the output controls
executed through function setup on the main unit of the balance.
The commands [O9] are data request commands issued to the balance.
※
Once any command from [O0] to [O9] is executed, the balance runs that function until another
command is entered. However, if the balance is switched off and on again, the output control is
reset to the initial function (function set value).
9.5.3 Response Output
(1) Response output format
Composed of five characters, including the terminators (CR = 0DH; LF = 0AH)
1 2 3 4 5
A1 A2 A3 CR LF
(2) Types of response outputs
A1 A2 A3
Code
Description
A 0 0
41H
30H
30H
Normal
end
E 0 1
45H
30H
31H
Command error
(Abnormal command received; other errors)
Содержание AJ-NT/DBNT Series
Страница 2: ......