19
5. Switching Function for Units of Measurement
5.1 Switching Units of Measurement
Pressing the Function key allows the user to switch the unit of measurement to [
g
], [ct] and so on.
Switching the units of measurement to be displayed
Press the Function key.
Each press advances through the units
of measurement to be displayed.
※
The balance is shipped from the
factory with the following default
setup: [
g
]
→
[ct]
→
[Pcs]
→
[
g
]
→
· · ·
5.2 Setup of Units of Measurement
When values [81.S.u.] to [85.S.u.] are entered prior to use, the desired unit of measurement to be
displayed can be chosen simply by pressing the Function key. For more information on the units
of measurement that can be set here, please refer to “4.2 Description of Functions” on page 17.
Example: To change the default factory settings to pound units, use [82.S.u.] .
1 Display the function item [82.S.u. 14]
Press and hold down the Function
key.
Release the key as [Func
]
appears on
the display.
The function setup mode is now
activated and the first item is displayed.
Press the Function key several more
times, until [82.S.u. 14] appears.
2 Setting the unit to a pound
Press the Zero key several times to
specify [82.S.u. 16].
Pcs
g
Содержание AJ-NT/DBNT Series