Cleaning the test port
It is a good idea to check that the optical connections are clean
and clean them if necessary before starting measurements. Even
very small dust particles on the end surfaces of the plugs or in
the test adapters can adversely affect the accuracy of the
measurement.
1.
Switch off the instrument.
2.
The most effective way is to use a specific MPO cleaner
(e.g. Senko, SmartCleaner MPO).
3.
Another way is to blow out the test adapter with clean
compressed air (available in spray cans, e.g. anti-dust spray).
Note:
Cover the optical connections with the dust cap whenever they
are not in use. This prevents them from getting dirty.
Cleaning the instrument
If the instrument gets dirty through use, you can clean it using a
soft cloth moistened with a mild solution of detergent.
WARNING
Dangerous voltage and invisible laser radiation
Maintenance or cleaning of the instrument while it is
connected or operating may damage the instrument or
injure you.
J
Make sure that the instrument is switched off and
disconnected from all power sources and optical radiation
sources before maintenance or cleaning.
NOTICE
Water and cleaning fluids
The instrument may be damaged or destroyed if water or
cleaning fluids penetrate it.
J
Make sure that water or cleaning fluids do not penetrate the
instrument.