Vetario T40M Скачать руководство пользователя страница 20

HD453 GB Issue 08 

 

20  
 

 

Not  suitable  for  use  in  the  presence  of  a  flammable  anaesthetic  mixture.  (The  incubator  is  not 
intended  to  be  used  in  a  location  where  flammable  anaesthetics  may  be  present.  This  would 
create a serious risk of fire or explosion). 

 

Before oxygen administration, disconnect any nebuliser and tube from the solution holder. (This 
is to prevent oxygen flowing into the nebuliser pump and creating a fire hazard). 

 

If  fitted,  disconnect  the  water  pump  control  lead  and  water  delivery  tube.  (The  pump  is  not 
intended for use during oxygen therapy and may create a fire hazard). 

 

Do  not  cover  the  intensive  care  unit.  (Blocking  of  ventilation  slots/holes  may  lead  to  oxygen 
enrichment  of  air  in  unintended  areas  leading  to  an  increased  risk  of  fire.  It  may  also  lead  to 
increased concentration of CO

2

 in the animal cabinet. Covering the unit may lead to fire hazard 

due to reduced dissipation of heat). 

 

Do  not  obstruct  ventilation  holes.  (Blocking  of  ventilation  slots/holes  may  lead  to  oxygen 
enrichment  of  air  in  unintended  areas  leading  to  an  increased  risk  of  fire.  It  may  also  lead  to 
increased concentration of CO

2

 in the animal cabinet). 

 

Do  not  exceed  1  lpm  (litre  per  minute)  oxygen  flow  rate.  (The  maximum  safe  long-term 
concentration of oxygen for canines and felines is achieved at this flow rate and increased flow 
may  lead  to  permanent  injury  or  death.  Greater  flow  rates  also  increase  the  risk  of  oxygen 
leaking in the event of a fault and could create a fire risk). 

 

The incubator contains no flow restriction or pressure regulator device. The oxygen supply must 
be  reliably  restricted  to  400kPa  (50psi).  Always  use  a  medical  grade  oxygen  pressure  regulator 
and  flow  meter  that  has  been  serviced  in  accordance  with  the  manufacturer’s  instructions. 
Failure of an external device may create a fire or pressure hazard or prevent effective treatment. 
It is assumed the operator will be trained in the safe and correct use of such devices. 

 

After  oxygen  administration,  animals  and  any  associated  cloth  materials  must  be  allowed 
sufficient  time  in  the  chamber  in  normal  air  to  allow  trapped  oxygen  to  disperse.  (Fabric  and 
similar materials that trap oxygen enriched air may burn vigorously if ignited by a spark or other 
source of ignition). Use only pure cotton materials to reduce the risk of static electricity sparks. 

 
Perform the following tests and safety checks before EVERY use for oxygen administration and fill in the 
Start-Up Check Sheet. 

11.1  Check the cabinet, door and hinges/lock for cracks, damage or loose parts that may allow oxygen 

gas to escape and create a fire hazard. Check for gaps between the upper and lower halves of the 
white cabinet. 

11.2  If  fitted,  disconnect  the  flexible  tube  from  the  nebuliser  solution  holder  to  prevent  any  flow  of 

oxygen to a nebuliser pump. This could create a fire hazard. 

11.3  Remove  the  water  pan  and  evaporating  block.  Disconnect  the  Brinsea  Advance  Humidity  Pump 

tube and electrical cable if fitted. Please note that it is not practical to use the pump during oxygen 
administration  due  to  capacity  restraints.  A  hydrated  oxygen  supply  is  commonly  used  and  helps 
humidify the animal chamber. 

11.4  Fit the oxygen hose barb to the inlet hole using the locknut supplied. Check for tightness if already 

in place, there must be no gap or oxygen gas could leak and create a fire hazard. The hose barb is 
suitable for hoses with a 6mm (1/4 inch) bore size. Always use a hose clamp device and ensure it 
is tightly fitted so that the hose cannot possibly fall off. 

 
 
 
 
 
 
 
 
 
 

 

Hose Barb. (Fit 
locknut on inside of 
cabinet). 

Содержание T40M

Страница 1: ...Introducing Your Patients 18 10 Internal Lighting 18 11 Oxygen Administration 19 12 Nebuliser 22 13 Cleaning and Disinfection 23 14 Safety inspection Maintenance and Calibration 25 15 Specification 33...

Страница 2: ...start unexpectedly Not suitable for use in the presence of a flammable anaesthetic mixture The incubator is not intended to be used in a location where flammable anaesthetics may be present This woul...

Страница 3: ...ng oxygen therapy and may create a fire hazard Always test the fan rotation and power failure alarms and check ventilation holes before using with oxygen gas If the alarm sounds during use disconnect...

Страница 4: ...ccordance with the instructions 3 Introduction The Vetario T40M and T50M provide the ideal veterinary intensive care environment for sick injured or postoperative animals and birds sensitive to hypoth...

Страница 5: ...air filters fitted 1 Cabinet base 1 Door 1 Hinge socket 1 Fastener and tool kit 1 Water pan 1 Evaporating block 1 Water funnel 1 Nebuliser solution holder 1 Nebuliser bracket 1 Hose barb for oxygen a...

Страница 6: ...nk on the right hand side of the home page to qualify for your free 3 year guarantee 5 Location and Assembly Ensure that the room temperature cannot drop on a cold night Ideally thermostatically contr...

Страница 7: ...the left as supplied but may be reversed to provide better access if required Fit the top hinge socket to the cabinet using the M5 x 16mm countersunk screws M5 flat washers and M5 dome nuts 5 3 5 4 P...

Страница 8: ...g the solution holder pipe in the slot in the bracket 5 9 Place the intensive care unit on a scratch and moisture resistant surface Worktop height is ideal Clear the area around the unit of any electr...

Страница 9: ...t up Check is to ensure the safety of the animal and the operator Failure to check each item could result in a hazard If at any stage the behaviour of the control system is not as stated then call Bri...

Страница 10: ...ucing animals or starting oxygen supply The unit is now ready for use See section 11 for further checks when using the unit for administration of oxygen If any stage of the Start Up Check results in a...

Страница 11: ...mains on the display to show this has happened Press to clear the indicator If the cause is not immediately apparent discontinue use and contact Brinsea Products 6 12 LOW TEMPERATURE ALARM DISPLAY If...

Страница 12: ...g DISCONTINUE USE AND CONTACT BRINSEA PRODUCTS Press the start button Does the heater control initiate and the green indicator illuminate Make sure the unit is not near a source of heat Wait for the a...

Страница 13: ...T INCUBATION TEMPERATURE DEFAULT 2 0 C 3 6 F LOW TEMPERATURE ALARM RANGE 1 0 5 0 C 1 8 9 0 F BELOW SET INCUBATION TEMPERATURE DEFAULT 3 0 C 5 4 F CELSIUS FAHRENHEIT DISPLAY SWITCHES ALL TEMPERATURE FI...

Страница 14: ...nce Sunlight can move round a room This could over heat the animal inside Pass Fail Ensure there are no electrical devices in the unit such as heat pads These can be a fire hazard if oxygen is being u...

Страница 15: ...ire in oxygen Pass Fail Check the oxygen supply hose for splits or cracks that could leak and create a fire risk Pass Fail Ensure the oxygen supply hose connection barb is tightly fitted and therefore...

Страница 16: ...multaneously to unlock the main menu Scroll to the C F option and press to select the C F display screen Press the button to select F or the button to select C Press to return to the Main Menu and the...

Страница 17: ...ce with the manufacturer s instructions in the water reservoir pan to inhibit bacterial build up It is recommended that the reservoir is topped up with solution daily to reduce dehydration This can be...

Страница 18: ...nimal to escape and lead to a consequent fall or other injury Monitor patients and device settings displays frequently so that incorrect operation or deterioration of patient condition is detected pro...

Страница 19: ...s are not introduced on a patient The incubator must only be used in a well ventilated room away from potential sources of heat or ignition NO SMOKING Do not use in a confined location the incubator n...

Страница 20: ...meter that has been serviced in accordance with the manufacturer s instructions Failure of an external device may create a fire or pressure hazard or prevent effective treatment It is assumed the oper...

Страница 21: ...intensive care unit for any reason as this may cause the unit to overheat or cause oxygen to collect around electrical parts This would create a serious fire hazard Ensure no items can fall onto the i...

Страница 22: ...equently for correct pressure and flow rate of oxygen and for correct incubation temperature Frequently check the vent is not obstructed 11 17 Under normal conditions the maximum sound level in the un...

Страница 23: ...ed if necessary Remove all three filters and gently hand wash in warm water then allow them to dry before use Filters need replacing every six months The exterior of the unit may be cleaned with a dam...

Страница 24: ...ate 13 9 ENSURE THE LED WIRES ARE NOT TOUCHING THE HEATING ELEMENT 13 10 Replace the metal cover so that the water pan slots are at the correct end Hold the cover in place and fit the 4 cap head screw...

Страница 25: ...R IN THE RESERVOIR PAN HUMIDITY PUMP CYCLE TIME ONLY APPLICABLE WHEN USED WITH THE BRINSEA ADVANCE HUMIDITY PUMP AMBIENT TEMPERATURE COMPENSATION AMBIENT TEMPERATURE USED IN HEATER CONTROL ALGORITHM O...

Страница 26: ...ever deliberately block ventilation holes or slots The on going safety of the incubator depends on correct servicing using exactly specified parts All servicing must be carried out only by a suitably...

Страница 27: ...able not crumbling or flaking by carefully pressing with the tip of a blunt object Contact Brinsea Products in the event the grommets or sealant need replacement due to aging Cracked or missing gromme...

Страница 28: ...es on the heater plate Ensure the LED power cables are clear of the heating element Replace the metal enclosure cover and tighten cap screws 14 8 Check electrical enclosure and cover for impact damage...

Страница 29: ...Measured value Triac TR2 shall be open circuit check T14 N neutral input terminal to T6 link by side of TX2 Measured value 14 10 Check safety critical components on the heater controller board BPL83 a...

Страница 30: ...necessary to temporarily fit the heater live terminal HL wire to the Test terminal 14 12 Replace control enclosure cover and perform earth continuity and electrical insulation tests Earth continuity M...

Страница 31: ...een LED s both OFF Record yes no Audible alarms OFF Record yes no 14 20 Press the start button Record the operation below incorrect function must be investigated before the unit is used Alarm sounds f...

Страница 32: ...ature to 36 0 C Allow at least an hour for the temperature to stabilise If adjustment is necessary press the OK and buttons simultaneously to access the calibration menu Select the first option CAL TE...

Страница 33: ...natively in some countries you may be able to return your product to your local retailer upon purchase of an equivalent new product Disposing of this product correctly will help save valuable resource...

Страница 34: ...of Issue 32 33 Buckingham Road Weston Industrial Estate Weston super Mare North Somerset BS24 9BG United Kingdom Declaration of Conformity We BRINSEA PRODUCTS LTD 32 33 Buckingham Road Weston Industr...

Отзывы: