Oxygen Administration
WARNING. RISK OF SERIOUS INJURY OR DEATH OF OPERATOR OR PATIENT:-
VETARIO TM INTENSIVE CARE UNITS HAVE BEEN SPECIFICALLY DESIGNED AND TESTED
FOR SAFE OPERATION WITH INCREASED OXYGEN CONCENTRATION OF AIR WITHIN THE
CABINET. ALL OPERATIVES MUST READ AND UNDERSTAND THE SAFETY INSTRUCTIONS.
Complete a Start-Up Check Sheet each time the unit is used for oxygen service.
Oxygen concentrations above normal atmospheric conditions create a greatly increased risk of
fire, including immediately outside the apparatus. Materials that do not burn in air may react and
burn vigorously in an oxygen enriched atmosphere.
No auxiliary equipment shall be placed in the incubator. Only use patient connections that are
designed for safe use in an oxygen enriched atmosphere. If in any doubt check with the
manufacturer of the auxiliary equipment. (Fire may be caused by sparking contacts or hot
surfaces in equipment not designed for oxygen service).
Even small quantities of flammable agents, such as ether and alcohol, left in the incubator can
cause fire in connection with oxygen. (Never use flammable cleaners or disinfectants and ensure
such agents are not introduced on a patient).
The incubator must only be used in a well-ventilated room away from potential sources of heat or
ignition. NO SMOKING. Do not use in a confined location, the incubator needs at least 30cm of
free space above and to both sides. (Oxygen concentration in the air around the incubator may
rise significantly if ventilation isn’t provided or the space around the incubator is confined. This
may create a fire hazard).
Remove all electrical devices (including the water pump) that are within 90cm of the incubator.
Place the incubator at least 90cm away from power sockets. Ensure there are no power sockets
or electrical items directly under the shelf or table supporting the incubator. (Oxygen gas may
sink through air and cause fire in other equipment).
Always test the fan rotation and power failure alarms and check ventilation holes before using
with oxygen gas. If the alarm sounds during use disconnect the oxygen supply, disconnect
power and open the door to ventilate the care unit. Do not re-use until the fault has been
investigated and rectified. (Failure of the ventilation fan or blocking of ventilation slots/holes may
lead to oxygen enrichment of air in unintended areas leading to an increased risk of fire. It may
also lead to increased concentration of CO
2
in the animal cabinet).