SLICE-QTC Four-Channel Temperature Controller
Product Manuals - https://www.vescent.com/manuals/
Fig. 24: Menu for setting absolute extrema and lock range
Locked
Range
ºC
Defines the nominal operating
temperature range for the plant. When the
actual temperature is within this window
(±value), the temperature error will be
displayed on a green background. When
the error in the temperature is outside this
window, the temperature error will be
displayed on a yellow background.
Maximum
Temp
ºC
Sets the maximum and minimum
operating temperatures of the plant. The
user will be prevented from adjusting the
setpoint temperature to outside this
range. If the actual temperature exceeds
this range, the temperature will be
displayed on a flashing red background.
Minimum
Temp
ºC
Shutdown
Time
s
Sets the delay between when the
temperature goes out of range and when
the SLICE-QTC shuts the current off.
Tab. 6: Warning and Max/Min
If your plant is exposed to the atmosphere, use an abundance of caution in setting T
min
below
the dew point for your ambient conditions! Water will condense on surfaces that are below
the dew point temperature.
Power Limits
The User can limit the control power available to individual channels. To adjust control capacity from
the Home screen, select CH X > Settings > Load Limits and the screen shown in
will appear.