Important Control Solution Information
Use only
control solutions.
Check the expiration date on the control solution vial. Do not use if ex-
pired.
Control solution, meter, and test strips should come to room temperature
(68~77ºF/20~25ºC) before testing.
Shake the vial, discard the first drop of control solution, and wipe off the
dispenser tip to ensure a good sample and an accurate result.
Use only for 90 days after first opening. Record the discard date (date
opened plus 90 days) on the control solution vial. Discard after 90
days.
Store the control solution tightly closed at temperatures below 86ºF
(0ºC). Do not refrigerate.
The control solution range printed on the test strip vial is for
control solution only. It is used to test meter and test
strip performance. It is not recommended range for your blood
glucose level.