background image

2

Go Direct Electrode Amplifier mV response is dictated by the amplifier (as
opposed to the electrode) and does not change over time, temperature, or with
the electrode being attached. Calibration of Go Direct ORP in mV by putting it
into different standards does not provide a better mV calibration. It provides a
worse calibration as it will assume the ORP electrode is perfect. Then the mV
reading will be off regardless of which electrode is connected.

Specifications

Response time

90% of final reading is one second in a buffer

Range

–1,000 mV to +1,000 mV

Typical accuracy

± 15 mV

Resolution

0.01 mV

USB specification

2.0

Wireless specification

Bluetooth 4.2

Maximum wireless range

30 m

Dimensions

EA: 8.5 cm × 3 cm × 1.75 cm

ORP: 15.5 cm long, 12 mm OD

Battery

300 mA Li-Poly

Battery life (single full
charge)

~24 hours

Battery life (long term)

~500 full charge cycles (several years depending
on usage)

Care and Maintenance

Short-term storage (up to 24 hours):

Place the electrode in pH 4 or pH 7 buffer

solution. It should never be stored in distilled water.

Long-term storage (more than 24 hours):

Store the electrode in a pH 4

buffer/KCl storage solution in the storage bottle. The ORP electrode is shipped
in this solution. Vernier sells 500 mL bottles of pH Storage Solution (order code
PH-SS), or you can prepare additional storage solution by adding 10 g of solid
potassium chloride (KCl) to 100 mL of pH 4 buffer solution. Vernier sells a pH
Buffer Capsule kit (PH-BUFCAP) that includes a buffer solution preservative.
Storing the electrode in this solution contributes to electrode longevity and
retains electrode response time when the unit is placed back into service.

If the electrode is inadvertently stored dry for a short period of time, immerse the
tip in the pH 4 buffer/KCl storage solution for a minimum of 8 hours prior to
use. If the readings are still not accurate after calibration or if the response is
slow, try shocking the sensor as described in the Troubleshooting section.

Connecting via Bluetooth

Ready to connect

Red LED flashes when sensor is awake and
ready to connect.

Connected

Green LED flashes when sensor is connected via
Bluetooth.

Connecting via USB

Connected and charging

Blue and Green LED are solid when sensor is
connected to Graphical Analysis via USB and
unit is charging. (Green LED is obscured by the
blue one.)

Connected, fully charged

Green LED solid when sensor is connected to
Graphical Analysis via USB and the unit is fully
charged.

Charging via USB,
connected via Bluetooth

Blue LED is solid and green LED is flashing, but
the green flashing LED looks white because it is
overwhelmed by the blue.

Identifying the Sensor

When two or more sensors are connected, the sensors can be identified by
tapping or clicking Identify in Sensor Information.

Using the Product

1. Remove the storage bottle from the electrode by unscrewing the lid and

removing the bottle and lid.

2. Thoroughly rinse the lower section of the probe, especially around the bulb-

shaped tip, using distilled or deionized water.

3. Connect the sensor following the steps in the Getting Started section of this

user manual.

4. When you are finished making measurements, rinse the electrode with

distilled water.

5. Slide the cap onto the electrode body, and then screw the cap onto the

storage bottle so the tip of the electrode is immersed in the storage solution.

Important:

Do not fully submerge the sensor. The BNC connection is not

waterproof.

Do not leave the electrode in acids or bases with concentrations greater than
1.0 M for periods longer than 5 minutes.

By default, the units displayed will be in mV. See

www.vernier.com/til/3984

to

change the displayed units.

Calibrating the Sensor

Calibration of Go Direct ORP is not supported. ORP readings are in mV. The

Отзывы: