Page | 6
except in case of emergency, as this will impair the effective operation of the
machine and may cause damage to the components.
1.4.1
Motor
1.1 KW; 3 Phase; 220-240v 50Hz/ 380-420v 50Hz; IP55; D90 frame size. Rating at
peak load 4.57 amps (nominal).
1.4.2
Invertor
AC invertor with either 110v or 230v supply voltage, 230v motor output and utilising
current monitoring.
1.4.3
Pump
0.27 kW; 24vdc; 50 cycles. 40 l/min at 5 lbs/in². Protected by a 5 amp fuse.
1.5
Water requirements
Inlet supply: Cold water
ONLY
with flow rate of 18 l/min (4 gals/min) nominal. Nominal
overall usage; 25 litres (+/- 10%) per cycle.
1.6
Drainage
A 2 inch/ 50mm P Trap compression fitting to be connected either directly, or via a swept
bend, to a standard 2in/50mm drain pipe with a 1 in 25 or suitable fall into a soil drain.
A 1 1/4 inch/32mm tank connector is supplied, which should be used to connect the
water tank warning pipe to a separate drain line, according to the relevant water
regulations.
A standard warning pipe installation kit is available. Contact Vernacare Customer Care
Department for details.